![HBK FS22SI Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/hbk/fs22si/fs22si_user-manual_2111679023.webp)
23
FS22SI
OPERATION
Important
Before using the provided LabVIEW
TM
examples on a PC, make sure the Regional and
Language options (in standards and formats) are defined as dot for decimal symbol,
comma for digit grouping symbol and dash for date separation.
Operational States
The FS22 - Industrial BraggMETER SI interrogator has six operational states:
The error state corresponds to a malfunction of the optoelec
tronic module of the interrogator. The system can only recover
from this state by resetting.
In the ready state, the FS22 - Industrial BraggMETER SI is ready
to start acquisition.
The free acquisition state enables the FS22 - Industrial
BraggMETER SI interrogator to perform single measurements on
a specified optical channel. The interrogator can be configured
in terms of threshold and the corresponding configuration can
be stored for further measurements.
The continuous acquisition state allows continuous wavelength
acquisition over all optical channels and also performs continu
ous OSA trace acquisition at 1 S/s.
The scheduled acquisition state allows periodic data acquisition
with a pre-defined period and number of samples while saving
data to the internal interrogator memory. Whenever the acquisi
tion period minus the number of samples (at 1 S/s rate) is higher
than 10 minutes, the interrogator will enter into sleep mode dur
ing inactive time intervals, to reduce power consumption. The
minimum period is 5 s and the number of samples (at 1 S/s)
cannot overcome the defined period.
The FS22 - Industrial BraggMETER SI interrogator starts on the
warming-up state that enables the settling of the operation
parameters and recall of the acquisition configuration.
0 Error
1 Ready
2 Free acquisition
3 Continuous
acquisition
4 Scheduled
acquisition
5 Warming-up
Содержание FS22SI
Страница 1: ...FS22SI Industrial BraggMETER SI ENGLISH User Manual...
Страница 18: ...FS22SI OPERATION 18 Fig 3 7 Local Area connection properties...
Страница 61: ...61 FS22SI BRAGGMONITOR SI SOFTWARE DETAILS Fig 6 2 Set Run as administrator automatically...
Страница 70: ...FS22SI BRAGGMONITOR SI USER INTERFACE 70 Fig 7 17 Select data file Fig 7 18 Text Import Wizard 1 3...
Страница 71: ...71 FS22SI BRAGGMONITOR SI USER INTERFACE Fig 7 19 Text Import Wizard 2 3 Fig 7 20 Text Import Wizard 3 3...
Страница 88: ...FS22SI BRAGGMONITOR SI USER INTERFACE 88 Fig 7 40 Update info window...
Страница 89: ...89 FS22SI BRAGGMONITOR SI USER INTERFACE...
Страница 90: ...HBK Hottinger Br el Kjaer www hbkworld com info hbkworld com A04249 10 E00 00...