![HBK FS22DI Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/hbk/fs22di/fs22di_user-manual_2111693034.webp)
FS22DI
REMOTE COMMUNICATIONS
34
R
Sampling
Rate
Integer. Valid Sampling Rates: 1000, 500, 200,
100, 50 (S/s).
N
Number
Integer.
HH:MM:SS
Time
Integer.
HH – hour
MM – minutes
SS – seconds
YYYY:MM:DD
Date
Integer.
YYYY – year
MM – month
DD – day
Q
NTP Quality
Integer.
0 for “Not acceptable”
1 for “Acceptable”.
5.1.2
Common Commands
:IDEN?
Queries the FS22 - Industrial BraggMETER DI interrogator
type and firmware revision. The return string is a list sepa
rated by
"
:
"
consisting of Manufacturer, Unit Name and Revi
sion, Number of Channels, Serial Number and Date.
Example:
Command
:IDEN?
Answer
:ACK:HBM FiberSensing:FS22DI
v2.0:08:046 840 200 188:20150202
Configuration
:STOR
Stores FS22 - Industrial BraggMETER DI interrogator acquisi
tion settings to the internal configuration file. The gain and
threshold values of each channel are stored. This command
can only be executed in the free acquisition state (
see section
Operational States for state description
).
Example:
Command
:STOR
Answer
:ACK
Identification
Store configuration