Each optical Interrogator can synchronize its internal clock with an NTP server. It is possi
ble to achieve accuracies of 1ms or higher, depending on whether or not a dedicated NTP
server is being used.
Further information about NTP can be found at
Measuring systems typology
A measurement system can be as simple as a single interrogator or a bit more complex
with combined interrogators with the same or different sampling rates, and interrogators
combined with other equipment.
For the usage of a single interrogator no special synchronization
is needed.
BraggMONITOR DI software can only operate one interrogator at
a time. If an NTP server is running on the same PC as the Brag
gMONITOR DI, the interrogator internal clock – hence Brag
gMONITOR DI – will sync.
If more than one interrogator is used, synchronization becomes
important and if an NTP server is running the equipment will
start synching as soon as a first communication from the server
is received.
Ethernet
Switch
NTP
NTP
NTP
NTP Server
Fig. 3.10
Synchronizing multiple interrogators
A hybrid measurement system is a system composed by differ
ent equipment types and technologies that can be operated with
different sampling rates. Taking the example of HBK families a
hybrid system could be a combination of Interrogators with
QuantumX
.
Single interrogator
Multiple
interrogators
Hybrid