Assembly Instructions M8086-EN 

 

M8086-EN   2020-02 

 

6/6

 

 

  

 

14.

  Check flow rate calibration: 

  After calibration, run the unit for at least 5 minutes at 40 °C in 

normal mode. 

  Close shut-off valve between feed and return lines. 

  Flow rate switches to 0 L/min and alarm indication 

‚Flow rate 

zero

’ is given. 

 

NOTE! 

The alarm message ‚Flow rate zero’ is delayed in 
time. 

  If this is not the case, repeat the flow rate calibration. 

  

 

  

 

ATTENTION! 

Risk of incorrect calibration! 

False calibrations can lead to faults witch the unit. 

Therefore: 
–   Check the calibrations. 

   

 

 

 

15.

  Switch the unit off by press the 

 key. 

 

 

 

 

16.

  Main switch off. 

 

 

 

 

17.

  Close the unit correctly again. 

  

Parts list 

       

 

 

Pos 

Description 

O/ID 

O/ID 

T2

4

6

2

9

-1

 

T2

4

6

2

9

-2

 

T2

6

3

2

6

-1

 

T2

6

3

2

6

-3

 

T2

5

2

4

2

-2

 

T2

5

8

6

5

-1

 

T2

5

8

6

5

-3

 

Pcs  Pcs  Pcs  Pcs  Pcs  Pcs  Pcs 

01 

Sonic transducer water up to 160 °C, 1 m 

T27045-1 

02 

Sonic transducer water up to 160 °C, 1,45 m 

T27045-3 

03 

Sonic transducer water up to 180 °C, 1 m 

T26310-1 

04 

Sonic transducer water up to 180 °C, 1,45 m 

T26310-3 

05 

Sonic transducer oil up to 200 °C, 1 m 

T26328-1 

06 

Sonic transducer oil up to 200 °C, 1 m 

T26082-1 

07 

Sonic transducer oil up to 250 °C, 1,45 m 

T26082-3 

08 

Socket 27x14x9 PEEK 

T26080 

09 

O-Ring FPM 17x2 mm 

T23523 

10 

O-Ring FFKM 17x2 mm 

T25429 

11 

Assembly instructions German 

M8086-DE 

12 

Assembly instructions English 

M8086-EN 

13 

Assembly instructions French 

M8086-FR 

 

 