Assembly Instructions M8086-EN
M8086-EN 2020-02
3/6
Fig. 1: Removing level indicator
Fig. 2: Pulling out level indicator
Only on T25865-1 and replacement sonic transducer BB 1
5.
To enable accessibility to the transducer (BB 1), the level
indicator (10) must be removed. Proceed as follows:
Undo screws on pipe (Label 1
Remove 6x screws (Label 2
Pull out level indicator (Label 3
Fig. 2) so that accessibility to
the transducer (BB 1) is ensured.
6.
Proceed as follows in order to remove the sound transducer
(BB 1, BB 2):
Remove the screws and mounting plate of the sound
transducer.
Remove sound transducer.
Remove O-ring.
Unplug cable of the sound transducer (BB 1, BB 2) on slot
X 53.1 or X 53.2.
7.
Proceed as follows to assemble the new sound transducer
(BB 1, BB 2)
Clean installation space of sound transducer, especially groove
for O-ring.
Insert the new O-ring in the groove.
Fit the sound transducer with the mounting plate and screws
carefully.
Transfer the cable labelling of the old sound transducer to the
new sound transducer.
Insert the sound transducer into the corresponding slot at the
X 53.1 or X 53.2.
0