Assembly Instructions M8098-EN 

 

M8098-EN   2018-03 

 

5/6

 

 

  

 

12.

  Reconnect mains plug and switch on main switch. 

 

 

13.

  Press the 

 key to switch on the unit and check it for leaks. 

   

 

14.

  Proceed as follows in order to calibrate the temperature 

sensor: 

 

When setting partameters set 

DFM recognition

 to the 

"integrated“ value (in 

Settings / Various)

, continue with Ste

15

. 

 

When setting parameters set 

DFM recognition

 

to the "modular“ 

value (in 

Settings / Various)

, continue with Step 

16

.

 

 

  

 

 

Fig. 10: Sensor calibration during 
integrated operation 

Integrated operation 

15. 

Temperature sensor main line ext 1..8 and temperature 
sensor return line ext. Proceed as follows in order to calibrate 
1..8 , if necessary, the temperature sensor: 

  With a constant deviation set parameter 

Sensor… offset

 under 

Service / Calibration / Temperature

  With a linear deviation, set parameter 

Sensor… asc.cor.

 under 

Service / Calibration / Temperature

 

 

 

Fig. 11: Sensor calibration during modular 
operation 

Operation Modular 

16. 

Proceed as follows, if necessary calibrate the temperature 
sensor for the main line and temperature sensor of the return 
line:  

 

Select the module affected e.g. "A1“ with the 

 or 

 

buttons. 

  With a constant deviation set parameter 

Sensor… offset

 under 

Service / Calibration / Temperature

  With a linear deviation, set parameter 

Sensor… asc.cor.

 under 

Service / Calibration / Temperature

 

 

17.

  Switch the unit off by press the 

 key. 

 

18.

  Main switch off. 

  

 

Содержание M8098

Страница 1: ...outconversion work on the unit Procedure DANGER Danger to life caused by electric current Touching conductive parts causes a direct danger to life Therefore For all work on the electrical system for maintenance cleaning or repair work disconnect from the mains or disconnect all phases of the external power supply and secure them against being switched on again Check unit is isolated from power sup...

Страница 2: ...cated with the symbol 2 Main switch off remove the plug from the mains and empty the unit 3 Remove control cable to ext flow rate meter Flow 5 4 Detach the hexagon screw and remove ext flow rate meter Flow 5 Fig 1 Front side of the evaluation unit Fig 2 Rear side of the evaluation unit 5 Proceed as follows in order to remove the cover of the evaluation unit Remove screws Label 1 Fig 1 Detach screw...

Страница 3: ...placement temperature sensor return line BT 2 x Disassemble screws Label 6 Fig 4 and remove cover 7 Replace temperature sensor main line BT 1 or return line BT 2 x while proceeding as follows Fig 5 Flow 5 side view from the right Disconnect temperature sensor BT 1 BT 2 x at the flow rate measuring board DFM 51 A 9 x Disassemble temperature sensor BT 1 BT 2 x Fig 5 Transfer label of temperature sen...

Страница 4: ...er to attach the covers Fig 8 Flow 5 with ball valve Fig 9 Flow 5 view from below Attach insulation and cover plate mount screws Label 5 Fig 8 Attach insulation and cover plate mount screws Label 4 Fig 8 If present install shut off valve OUT 1 and 4 Label 3 Fig 8 Only in the event of a replacement temperature sensor return line BT 2 x Attach cover plate and mount screws Label 6 Fig 9 10 Connect co...

Страница 5: ... sensor return line ext Proceed as follows in order to calibrate 1 8 if necessary the temperature sensor With a constant deviation set parameter Sensor offset under Service Calibration Temperature With a linear deviation set parameter Sensor asc cor under Service Calibration Temperature Fig 11 Sensor calibration during modular operation Operation Modular 16 Proceed as follows if necessary calibrat...

Страница 6: ... EN M8098 EN 2018 03 6 6 Parts list Pos Description O ID Pcs 01 Heat sensor Pt 1000 0 53 m T24888 1 02 Assembly instructions German M8099 DE 1 03 Assembly instructions English M8099 EN 1 04 Assembly instructions French M8099 FR 1 ...

Отзывы: