6

Warning 

 

The operators who will do the endoscopic surgery with medical pressurizer must be 

trained and qualified. Forbid unqualified personnel to use this equipment. 

 This device must be installed, used, operated and maintained by or under supervision 

the person who is properly trained and qualified. 

 It should has reliable grounding with the power supply socket to control unit. The 

power grid connected to the operating room must be grounded reliably, or doctors and 
patients would result in injury. 

Note 

 Keep the User’s Manual in a place easily accessible at all times for persons operating 

and maintaining the equipment.   

 The manufacturer is only responsible of the safety, reliability and performance of the 

device: 

               1) Electrical installation should be done in accordance with the provisions described in 

this user manual. 

2) This device should be installed, used, operated and maintained in accordance with 

this user manual. 

 Please do not use this device if it is suspected to have problems. Please contact 

HAWK designated service providers and check the device before use. If any deficiency 
is suspected or confirmed in the device, do not use the device until it has been checked 
by the authorized service. 

 Do not disassemble other parts by unauthorized person except the parts permitted by 

this operation manual. The maintenance of this device should be done in accordance 
with this user manual.Any additional maintenance and modification should be done only 
by HAWK service technicians or authorized personnel. 

 Do not modify this device without any authorization. The manufacturer is not liable for 

damage caused by unauthorized tampering with the device(s). Such tampering will make 
customer forfeit any rights to claim under warranty. 

 This equipment may only be used together with accessories supplied by HAWK. If the 

customer makes use of other brand accessories, use them only if their safe usability 
under technical safety aspects has been proved and confirmed by HAWK or the 
manufacturer of the accessory. 

Before insert plug into socket,it should check the plug is match the socket, do not force 

to insert    the plug into the socket. Do not pull the cable when turn off the power supply, 
it should hold the plug and pull it out . 

 Do not force cable connections. If a cable does not connect easily, be sure that the 

connector (plug) is appropriate for the receptacle(socket). If you cause any damage to a 
cable connector(s) or receptacle(s), let the damage(s) be repaired by an authorized 
service technician. 

 Don’t plug the AC power cord into the outlet before the connection between devices of 

the equipment is completed which can generate the defect. 

 Before every operation, visually check the equipment for exterior mechanical 

damaged(s) and for proper function. 

Содержание DR060380B-2

Страница 1: ...1 VIDEO CYSTOSCOPE M o d e l D R 0 6 0 3 8 0 B 2 USER MANUAL MANUAL P N IA MN 213 3 Version 3 1 Hangzhou Hawk Optical Electronic Instruments Co Ltd...

Страница 2: ...ments We also welcome any comments you may have concerning about our equipment Your comments and questions will receive immediate attention You can contact us via phone mail Fax and E mail hawk chinah...

Страница 3: ...Class I BF Part 3 According to IEC 60529 2001 02 The protection level of monitor is IPX0 The protection level of endoscope is IPX7 2 Certification The general electrical medical system is CE ISO9002...

Страница 4: ...patibility 10 1 12 WEEE 13 2 Products description 13 2 1 Products list 13 2 2 Intended use 13 2 3 Scope of use 13 2 4 Feature 14 2 5 Electrical connections 14 3 Instruction of product 14 3 1 3 5 LCD 1...

Страница 5: ...You must following the proper steps in training yourself for use of this equipment Otherwise there may be detrimental consequence to patients or the operator Furthermore you should carefully study the...

Страница 6: ...permitted by this operation manual The maintenance of this device should be done in accordance with this user manual Any additional maintenance and modification should be done only by HAWK service tec...

Страница 7: ...ns Where device is exposed to chemical material or explosive gas Be careful not to be inserted dust especially metal Where the humidity is extremely high or there is a ventilation problem Be careful n...

Страница 8: ...g term use the endoscope may be damaged please be sure to regularly check Even a slight abnormality was found please be cautious careful examination Be sure not to disassemble or modify this videoscop...

Страница 9: ...and Symbol Label Icon Meaning AC Power Power supply Turn on Main power Turn off Main Power Type BF Applied Part Note Warning Fuse Manufacture Date Manufacturer and Address European Union representativ...

Страница 10: ...d below The customer or the user of the Irrigation Aspiration Unit should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions CISP...

Страница 11: ...al of hospital environment lf the user of the Ele Pneumatic Lithotriptor image intensifier requires continued operation during power mains interruptions it is recommended that the Ele Pneumatic Lithot...

Страница 12: ...a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and landmobile radios amateur radio AM and FM radio bradcast and TV broadcast cannot be predicted...

Страница 13: ...this equipment like domestic waste it should be divided and disposaled You can contact local waste disposal collecting department to get detailed information Illegal disposal of waste will be subject...

Страница 14: ...be used 2 4 4 It has fixed focus large depth of field wide angle and get clean image within the range of 3 50mm Note This equipment is according to the requirement of Application Part of Type BF it po...

Страница 15: ...Increase button 11 Back switch button 12 Scope Lock 13 MENU 14 Video button 15 Photo button 16 Power Indicator 17 Charging Indicator 3 2 8 LCD 18 Housing 19 AV output connector 20 USB connector 21 SD...

Страница 16: ...ut connector charge the monitor 7 USD connector transfer image to computer 8 SD Slot insert SD card 9 Decrease button Decrease the parameter and up the menu review the picture 10 Increase button Incre...

Страница 17: ...ture and confirm in menu 31 Power switch l ON O OFF Power lamp lights up when the power switch turned on 32 Power input connector charge the monitor 33 Charging indicator Indicator on means charging i...

Страница 18: ...e video files will be saved automatically The movie resolution can be chosen in the menu as AVI format NOTE 10 minutes 20 minutes 30 minutes is supported when capturing video when the capacity of sing...

Страница 19: ...rightness Language press and button to select then press photo confirm button to confirm 1 Video Time 10Mins 20Mins 30Mins can be selected 2 Storage Flash SD Card we recommend you store photos And vid...

Страница 20: ...ding to the date a file named according to the time 3 5 10 Delete Files Press Menu button to enter into delete menu when single file is displayed 3 options can be selected delete all of this current a...

Страница 21: ...deo can be taken SD card cannot be unplugged when it is in the working condition of videoscope otherwise the memory files will be damaged or even damage the SD card When SD card is inserted into SD ca...

Страница 22: ...heck the bending part can be bent successful and also can be reach the maximum bending angle by controling the bending wheel 4 2 4 Checking other parts 1 Check if there is any damage at the flexible t...

Страница 23: ...when operating CCD camera or it will impair the image quality 4 3 OPERATION METHOD Warning The videoscope operator should study the various clinical factors involved in the process and to the correct...

Страница 24: ...perator may hurt in order to prevent injury it should not take long observation time of videoscope tip near the mucosa or mucous membrane If the device in question is running the power switch must be...

Страница 25: ...he endoscope thus it is impossible to test leakage Please do not use the device and contact with us See picture 11 After leakage please release the pressure and take off the leakage measure if not may...

Страница 26: ...before you call for service If the problems still exit turn off the device and move monitor from the videoscope and contact local service technician or appointed HAWK service facility 5 3 1 Monitor Ph...

Страница 27: ...hand this device to HAWK service and the device must conform to the following requirements It should clean and sterilize the device before packing it This device is a precision device please use stro...

Страница 28: ...on part 5 7mm Instruments channel 2 2mm Working length 380mm Depth 3 50mm Bending angle Up 210 down120 Running environment Temperature 10 to 40 Humidity 30 75 Pressure 700 1060hPa Storage and transpor...