changes to white, and immediately release the button. 

4.1.5 Please aim the videoscope at an object (such as your hand), if there is abnormal dot on the 

monitor screen, blurred images or striped images. If the image is not clear, check the cable connection, 

or wipe the objective lens with clean gauze moistened with alcohol. 

4.2 Checking before Operation   

4.2.1Inspect and confirm each connection.   

4.2.2 Carefully wipe the insertion part, the bending part and the distal with alcohol cotton. 

4.2.3 Check endoscopes and accessories: Before each clinical use, endoscopes and their accessories 

should be checked in accordance with the following procedure. If you notice a slight defect insertion 

tube surface, the imaging blur; or when other abnormalities, do not use vidoscope, contact with the 

company promptly. 

4.2.3.1 Check the insertion tube 

a) Check if the surface of insertion tube is not smooth, damage and other defects. 

b) Chec if there are any protrusion, internal loose, rough surfaces, sharp edges and other defects. 

4.2.3.2 Check bending part 

Check if there is any damage and other defects in the surface, and check the bending part can be bent 

successful and also can be reach the maximum bending angle by controling the bending wheel. 

4.2.4 Checking other parts 

1) Check if there is any damage at the flexible tube of light guide part (such as cracks, distortion or 

flattened). 

2) Please be sure the monitor in right condition 

Warning 

 

Before each use, the outer surface of the portions of the ENDOSCOPES which are 

intended to be inserted into a PATIENT should be checked to ensure there are no 
unintended rough surfaces, sharp edges or protrusions which may cause HARM. 

 If any deficiency is suspected or confirmed in the device, do not use the device until it 

has been checked by the authorized se. 

 

Before each use or after a change of viewing modes / settings, the OPERATOR 

should check to ensure the view observed through the ENDOSCOPE provides a live 
image (rather than a stored one) and has the correct image orientation. 

 

INTERCONNECTIONS CONDITIONS require the APPLIED PARTS of other ME 

EQUIPMENT used within the CONFIGURATION FOR ENDOSCOPIC APPLICATION to 
be TYPE BF APPLIED PARTS. And the application of CF part of the equipment can not 
be used together, otherwise it will cause the application part of CF equipment leakage 
current exceeds the standard. 

 

Before each use, Check equipment accessories are used correctly. 

Содержание DR060380B-2

Страница 1: ...1 VIDEO CYSTOSCOPE M o d e l D R 0 6 0 3 8 0 B 2 USER MANUAL MANUAL P N IA MN 213 3 Version 3 1 Hangzhou Hawk Optical Electronic Instruments Co Ltd...

Страница 2: ...ments We also welcome any comments you may have concerning about our equipment Your comments and questions will receive immediate attention You can contact us via phone mail Fax and E mail hawk chinah...

Страница 3: ...Class I BF Part 3 According to IEC 60529 2001 02 The protection level of monitor is IPX0 The protection level of endoscope is IPX7 2 Certification The general electrical medical system is CE ISO9002...

Страница 4: ...patibility 10 1 12 WEEE 13 2 Products description 13 2 1 Products list 13 2 2 Intended use 13 2 3 Scope of use 13 2 4 Feature 14 2 5 Electrical connections 14 3 Instruction of product 14 3 1 3 5 LCD 1...

Страница 5: ...You must following the proper steps in training yourself for use of this equipment Otherwise there may be detrimental consequence to patients or the operator Furthermore you should carefully study the...

Страница 6: ...permitted by this operation manual The maintenance of this device should be done in accordance with this user manual Any additional maintenance and modification should be done only by HAWK service tec...

Страница 7: ...ns Where device is exposed to chemical material or explosive gas Be careful not to be inserted dust especially metal Where the humidity is extremely high or there is a ventilation problem Be careful n...

Страница 8: ...g term use the endoscope may be damaged please be sure to regularly check Even a slight abnormality was found please be cautious careful examination Be sure not to disassemble or modify this videoscop...

Страница 9: ...and Symbol Label Icon Meaning AC Power Power supply Turn on Main power Turn off Main Power Type BF Applied Part Note Warning Fuse Manufacture Date Manufacturer and Address European Union representativ...

Страница 10: ...d below The customer or the user of the Irrigation Aspiration Unit should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions CISP...

Страница 11: ...al of hospital environment lf the user of the Ele Pneumatic Lithotriptor image intensifier requires continued operation during power mains interruptions it is recommended that the Ele Pneumatic Lithot...

Страница 12: ...a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and landmobile radios amateur radio AM and FM radio bradcast and TV broadcast cannot be predicted...

Страница 13: ...this equipment like domestic waste it should be divided and disposaled You can contact local waste disposal collecting department to get detailed information Illegal disposal of waste will be subject...

Страница 14: ...be used 2 4 4 It has fixed focus large depth of field wide angle and get clean image within the range of 3 50mm Note This equipment is according to the requirement of Application Part of Type BF it po...

Страница 15: ...Increase button 11 Back switch button 12 Scope Lock 13 MENU 14 Video button 15 Photo button 16 Power Indicator 17 Charging Indicator 3 2 8 LCD 18 Housing 19 AV output connector 20 USB connector 21 SD...

Страница 16: ...ut connector charge the monitor 7 USD connector transfer image to computer 8 SD Slot insert SD card 9 Decrease button Decrease the parameter and up the menu review the picture 10 Increase button Incre...

Страница 17: ...ture and confirm in menu 31 Power switch l ON O OFF Power lamp lights up when the power switch turned on 32 Power input connector charge the monitor 33 Charging indicator Indicator on means charging i...

Страница 18: ...e video files will be saved automatically The movie resolution can be chosen in the menu as AVI format NOTE 10 minutes 20 minutes 30 minutes is supported when capturing video when the capacity of sing...

Страница 19: ...rightness Language press and button to select then press photo confirm button to confirm 1 Video Time 10Mins 20Mins 30Mins can be selected 2 Storage Flash SD Card we recommend you store photos And vid...

Страница 20: ...ding to the date a file named according to the time 3 5 10 Delete Files Press Menu button to enter into delete menu when single file is displayed 3 options can be selected delete all of this current a...

Страница 21: ...deo can be taken SD card cannot be unplugged when it is in the working condition of videoscope otherwise the memory files will be damaged or even damage the SD card When SD card is inserted into SD ca...

Страница 22: ...heck the bending part can be bent successful and also can be reach the maximum bending angle by controling the bending wheel 4 2 4 Checking other parts 1 Check if there is any damage at the flexible t...

Страница 23: ...when operating CCD camera or it will impair the image quality 4 3 OPERATION METHOD Warning The videoscope operator should study the various clinical factors involved in the process and to the correct...

Страница 24: ...perator may hurt in order to prevent injury it should not take long observation time of videoscope tip near the mucosa or mucous membrane If the device in question is running the power switch must be...

Страница 25: ...he endoscope thus it is impossible to test leakage Please do not use the device and contact with us See picture 11 After leakage please release the pressure and take off the leakage measure if not may...

Страница 26: ...before you call for service If the problems still exit turn off the device and move monitor from the videoscope and contact local service technician or appointed HAWK service facility 5 3 1 Monitor Ph...

Страница 27: ...hand this device to HAWK service and the device must conform to the following requirements It should clean and sterilize the device before packing it This device is a precision device please use stro...

Страница 28: ...on part 5 7mm Instruments channel 2 2mm Working length 380mm Depth 3 50mm Bending angle Up 210 down120 Running environment Temperature 10 to 40 Humidity 30 75 Pressure 700 1060hPa Storage and transpor...

Отзывы: