9.7 RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE
RF COMMUNICATIONS EQUIPMENT AND HAUSTED ESC2EYEST PROCEDURE CHAIR
1
The Hausted ESC2EYEST Procedure Chair is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the Hausted ESC2EYEST Procedure Chair
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the Hausted ESC2EYEST Procedure Chair as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter (W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.2√P
80 MHz to 800 MHz
d = 1.2√P
800 MHz to 2.7 GHz
d = 2.3√P
0,01
0.12
0.12
0.23
0,1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
9.8 NOTES TO SECTIONS 9.2 - 9.7
1. 60601-1-2 © IEC:2014
9.9 CERTIFICATIONS
STANDARD
Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And
Essential Performance [AAMI ES60601-1:2005+A1]
Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And
Essential Performance [CSA C22.2#60601-1-2014 Ed:3]
Medical Electrical Equipment - Part 2-46: Particular Requirements For The Basic
Safety And Essential Performance Of Operating Tables [IEC 60601-2-46:2016]
Medical Electrical Equipment - Part 1-6: General Requirements For Basic Safety And
Essential Performance - Collateral Standard: Usability [IEC 60601-1-6:2010 Ed.3+A1]
PRODUCT
Electrical Surgical Chair
BRAND NAME Hausted by Graham-Field
MODEL
ESC2EYEST
EQUIPMENT
DESIGNATION
ESC2EYEST to IEC 60601-1-2:2014 Class A for Emissions, Immunity for Professional
Healthcare Facility Environment
EN 60601-1-2:2015 / IEC 60601-1-2 [ed.4] - 2014 Clause 5 ME Equipment and ME
Systems identification, marking and documents