5 Operate
16
5.2
Using the K6 signal socket for the operation
Connecting the signal line K6 (accessory) to the K6 signals socket is
mandatory.
1. Use the power switch to turn on the power pack.
• To verify this, the green power switch is illuminated.
• The HV output voltage is indicated on the HV display of the
power pack.
• The power pack is running.
2. If an error occurs, pins 3 and 4 have continuity at the K6 signal
socket and the HV is switched off.
• The error lamp is blinking.
3. The HV can be switch on again by proceeding with a reset.
• To execute the external reset, close the potential-free normally
open contact (> 0.5 s) briefly.
• For a manual reset, simply press the reset button.
NOTE:
An error indication over the K6 signalling socket will indicate a fault.
The discharging power pack can be reset by triggering the external
reset. If the defect persists, refer to the following chapter
“Troubleshooting”. Refer page 19.