ElectroPrep System
Publication 9511-073 REV-1.0
Using the ElectroPrep Unit
Primary Link Chamber
Secondary Link
Dialysis Chamber
ElectroPrep Membrane
Open End Cap
Key
Open End Caps:
Lock membranes
into position.
Membranes:
Facilitate selective separation
of biological samples from 500
to 300,000 Da.
Dialysis Chamber:
The major receptacle for
sample or dialyzed product.
Secondary Link Chamber:
Can be connected to a
Primary Link Chamber, and
joined with an Open End Cap
at the other end. Membranes
of different MCWOs may be
used at junctions of these
chambers and caps.
Primary Link Chamber:
Can be connected to a
Dialysis Chamber, and
joined with a Secondary Link
Chamber or Open End Cap at
the other end. Membranes
of different MCWOs may be
used at junctions of these
chambers and caps for size
fractionation.
How to Select Your Chamber and Membrane Configuration
1. Decide upon your application, e.g., electrodialysis, electroelution, electrofiltration,
electroconcentration, electroseparation.
2. Select a Dialysis Chamber able to hold the desired sample volume (50 to 1500 µl.)
Note that two Dialysis Chambers can be joined with a Union (with or without
membranes added between the Union and Chambers) to increase sample
volume (up to 600 or 3500 µl).