DOUBLE LOOP CONTROL
12 of 17
SETTING THE LOW
TEMPERATURE LIMIT
For routine operation, the low temperature limit for
thermocouple 1 (TC1) and thermocouple 2 (TC2) should
be set to -200ºC, as follows:
•
Make sure that the units have been set to
Centigrade, as described above.
•
Simultaneously press both the Up-arrow
S
and
the Down-arrow
T
keys for approximately 3
seconds until the Input Menu prompt
InPt
appears
in the upper display and the Setup Menu prompt
SEt
appears in the lower display:
InPt / SEt
•
Press
the
MODE
key twice.
rL1
will appear in the
lower display.
•
Use the Down-arrow
T
key to set the upper
display to
-200
.
•
Press the MODE key five more times.
rL2
will
appear in the lower display.
•
Use the Down-arrow
T
key to set the upper
display to
-200
.
•
Press
the
DSPY
key to return to the main display
format.
SETTING THE HIGH
TEMPERATURE LIMIT
For routine operation, the low temperature limit for
thermocouple 1 (TC1) and thermocouple 2 (TC2) should
be set to
750ºC
, as follows:
•
Make sure that the units have been set to
Centigrade, as described above.
•
Simultaneously press both the Up-arrow
S
and
the Down-arrow
T
keys for approximately 3
seconds until the Input Menu prompt
InPt
appears
in the upper display and the Setup Menu prompt
SEt
appears in the lower display:
InPt / SEt
•
Press
the
MODE
key three times.
rH1
will appear
in the lower display.
•
Use
the
Up-arrow
S
and/or Down-arrow
T
keys
to set the upper display to
750
.
•
Press
the
MODE
key five more times.
rH2
will
appear in the lower display.
•
Use
the
Up-arrow
S
and/or Down-arrow
T
keys
to set the upper display to
750
.
•
Press
the
DSPY
key to return to the main display