SINGLE LOOP CONTROL
8 of 17
SETTING THE HIGH
POWER LIMIT
For routine operation, the high power limit should be set to
an initial value of 75% for heating to temperatures above
200ºC or to 35% for temperatures up to 200ºC.
•
Simultaneously press both the Up-arrow
S
and
the Down-arrow
T
keys for approximately 3
seconds until the Input Menu prompt
InPt
appears
in the upper display and the Setup Menu prompt
SEt
appears in the lower display:
InPt / Set
•
Press the Up-arrow
S
key twice.
9LbL
appears
in the upper display:
9LbL/ SEt
•
Press
the
MODE
key nine times.
HiP
will appear
in the lower display.
•
Use the Down-arrow
T
key to set the high power
limit, shown in the upper display, to
35
or
75
.
•
Press
the
DSPY
key to return to the main display
format.
Notes
For operation of the HVC-DRP-4 and HVC-VUV-4 diffuse
reflection reaction chambers at or near their high
temperature limit, set the high power limit to 92.
The high power limit controls the percentage of full power
that is sent out to the heaters. The optimum setting
depends on the set point temperature and on the heated
device. If this high power limit is set too high and the set
point is close to the high temperature limit of the heated
device, the heater(s) may heat too quickly and burn out. If
the high power limit is set too low, the heated device may
heat too slowly and may not reach the set point. Adjust
this setting as needed from the recommended initial
settings to achieve the desired heating profile.
SETTING THE AUTO-TUNE
SET POINT
•
Simultaneously press both the Up-arrow
S
and
the Down-arrow
T
keys for approximately 3
seconds until the Input Menu prompt
InPt
appears
in the upper display and the Setup Menu prompt
SEt
appears in the lower display:
InPt / Set
•
Press the Up-arrow
S
key twice.
9LbL
appears
in the upper display:
9LbL/ SEt
•
Press
the
MODE
key ten times.
AtSP
will appear
in the lower display.
•
Use
the
Up-arrow
S
and/or the Down-arrow
T
keys to set the set point percentage, shown in the
upper display, to
90
.
•
Press
the
DSPY
key to return to the main display
format.