38
Cause #3:
Instrument or floor scale improperly grounded (i.e.
one or both not connected to earth ground).
Remedy #3:
Make sure that the weigh instrument and the floor
scale platform cover are both connected to earth
ground.
Cause #4:
High static electricity present.
Remedy #4:
Install a ground strap from the floor scale platform
cover to earth ground.
WARNING
DO NOT CONNECT THE GROUND STRAP
NEAR OR ON THE LOAD SENSORS. THEY
ARE GROUNDED THROUGH THE SUMMING
JUNCTION BOX. TO DO SO WILL DAMAGE
THE LOAD SENSORS.
INSTRUMENTATION
INTERFACE CABLE
COLOR CODES
-EXCITATION
RED
+SENSE
BLUE
+SIGNAL
GREEN
-SIGNAL
WHITE
-SENSE
BROWN
-EXCITATION
BLACK
+C2
GREY
-C2
VIOLET
SHIELD
ORANGE