HI 1769-WS/HI 1769-2WS WEIGH SCALE MODULE
3-14
FIG. 3-15 RTA SCHEMATIC HI 1769-2WS
EMI Suppression
Core Installation
(Prt. #2547-0013)
For CE requirements you will need to install an EMI
suppression core around the multi-strand portion of
the RTA cable. (See Fig. 3-12 & 3-14)
NOTE:
Install one suppression core for the single channel
model and two (2) suppression cores for the dual
channel model.
Step 1.
There should be enough of the individual
strands of wire exposed to install the sup-
pression core. If there is not enough room,
remove enough of the cable cover until
you can place the suppression core around
all the wire strands.
Step 2.
Place the suppression core as close to the
module door as possible and still be able to
close the door.
Step 3.
To open the suppression core place a small
slotted screwdriver behind the latch and
pry the latch away from the body of the
core until it clears both catches. (See Fig.
3-16)
Содержание HI 1769-2WS
Страница 14: ...HI 1769 WS HI 1769 2WS WEIGH SCALE MODULE 1 6 ...
Страница 27: ...Chapter 3 Installation 3 9 FIG 3 7 RAIL FASTENERS IN THE CLOSED POSITION ...
Страница 60: ...HI 1769 WS HI 1769 2WS WEIGH SCALE MODULE 4 26 ...
Страница 74: ...HI 1769 WS HI 1769 2WS MANUAL 6 8 ...
Страница 82: ...HI 1769 WS HI 1769 2WS MANUAL Weighing System Tests 1 4 wrist strap grounding device 3 2 ...