Revised 2020.1 6.12.2020
MODELS N1035P & N2085P
HOME CARE MANUAL STOCKTON BED
CAREGIVER INFORMATION &ASSEMBLY INSTRUCTIONS
The nations #1 manufacturer of hospital cribs
Страница 1: ...Revised 2020 1 6 12 2020 MODELS N1035P N2085P HOME CARE MANUAL STOCKTON BED CAREGIVER INFORMATION ASSEMBLY INSTRUCTIONS The nations 1 manufacturer of hospital cribs ...
Страница 2: ...nd beds for up to 12 years from the date of manufacture as long as we still make the part or can obtain it from our vendors HARD Manufacturing cribs and beds can be cleaned with an anti bacterial all purpose cleaner of the parent caregiver choice but it must be a non abrasive cleaner Anything containing bleach alcohol or ammonia must be avoided The crib should be dried thoroughly with a soft cloth...
Страница 3: ...ase until that time Letting the red tips of the trigger mechanism drag up and down the slide rods by releasing the trigger mechanism too early will cause the trigger assembly to wear prematurely and break The side rails trigger mechanism in the side rail must lock into place to hold up the rail securely and safely The red tips of the trigger assembly must be completely inserted into the notches in...
Страница 4: ...crank handle will continue to turn after 17 revolutions however it will become noticeably more difficult Continued cranking will not raise the head of the bed any further it will only put stress on the crank handle components and eventually cause them to break CRANK HANDLE USAGE All of the Stockton or Monroe Beds for home care use made by Hard Manufacturing have head and knee foot elevation as well ...
Страница 5: ...sed approximtely 10 to help prevent caregiver fatigue The 10 maximum is reached with 39 turns of the Hi Lo crank handle The Foot Crank Handle The foot crank handle is located on the far right of the foot end of your bed The foot crank handle when not in use should be tucked in This helps to avoid damage from it hitting or getting caught on other objects when the bed is being moved To use the foot ...
Страница 6: ......
Страница 7: ...the 4 casters to the bed with the 4 bolts contained in the box 4 Roll the bottom frame under the top frame as shown in the picture below 5 a Remove the 8 bolts and nuts located on both sides of the bed frame b Connect the bottom and top frame by installing and tightening the 8 bolts which were removed in a Instructions for the Assembly of N1035 N2085 Stockton Bed Manual ...
Страница 8: ...e head and foot ends to the spring frame Tap the corner lock on the end with a mallet and wood block until they are firmly seated in the corner lock on the spring frame 9 Make sure the hole of the head and foot ends align with the spring frame hole Then using the locking screw which was removed in step 7 screw and tighten the two components together as shown below ...
Страница 9: ...he purpose of shipping Remove Bolts Attach the bolts again after connecting to pivot assembly 11 Attach the slide rod extensions to both ends of the side frame with the plastic stop facing towards red trigger 12 Install the spring system to the slide rod extensions a Place one washer 1 x 0 545 x 0 125 into stem of slide rod b Next attach the bumper spring c Attach the final washer 1 x 0 545 x 0 12...
Страница 10: ...nd of the side frame to engage the square bushing into the upper crib eye b Place washer on top of crib eye and then tighten the locking screws at the top of each slide rod extension c Repeat procedure for all 4 slide rod extensions Total of 4 screws 4 washers Ensure side frame is under crib eye Tighten side frame to bed using bolt and spacer Completion of step 15 Assembled N1035 N2085 Stockton Be...
Страница 11: ...TOCKTON OR MONROE YOUTH BEDS The nations 1 manufacturer of hospital cribs INSTALLATION USAGE INSTRUCTIONS FOR HEAVY DUTY BUMPER PADS Revised 2020 1 3 25 2020 1035BP fits a 36 x 72 Bed 2085BP fits a 36 x 83 Bed ...
Страница 12: ...he mattress in place Place 1 square shaped pad against the inside surface of the headboard with the 2 white strings hanging down to the corner brackets of the spring assembly The corner brackets are the areas where the mattress platform attaches to the legs of the bed Put the vinyl flap with the Velcro over the top edge of the headboard and affix it to the outside face of the headboard This secures t...
Страница 13: ...the side rail Be careful not to cover the 2 openings for the fingertip trigger assembly mechanisms With the side rail in highest position and locked in place place one of the pads along the inside face of the side rail with the elastic straps and white strings towards the floor IMPORTANT The triangle shaped area on the pad the area that doesn t have any padding must go towards the head end of the be...
Страница 14: ...NG THESE STEPS Lift the 3 vinyl flaps along the entire length of the top of the side rail and then lower the side rail Pull the entire bumper pad towards you and allow it to fall outside the bed while staying attached to the bed at the bottom RAISING THE SIDERAILS WITH THE BUMPER PADS ON Lift the pad up and reach down to raise the side rail When the side rail is raised and locked into place reattac...
Страница 15: ...ospital cribs All bolts and other fasteners must be securely tightened and maintained when crib is put in service Before each usage or assembly inspect crib for damaged hardware loose joints missing parts or sharp edges CAUTION SHEET ...
Страница 16: ...replacement of the mattress Expected Service Life of HARD Manufacturing Cribs or Beds is 9 years Five 5 year warranty HARD Manufacturing will replace at no cost to the purchaser any expendable component such as but not limited to mattresses trigger assemblies slide rods or slide rod components crank handles Johns Hopkins handles vinyl curtains curtain rods or IV components found to have a defect i...