To disassemble the crib and replace a side rail, end rail or a slide rod:

**We recommend this be done with 2 people**

2. Push down on both slide rods until they are clear of the upper crib eye, then lean the
entire side assembly outwards. Lift the assembly up and o

  of the lower crib eyes

being careful not to lose any of the small parts at the bottom of the slide rods as the spring,
washers and spacer will fall o

 .

3. Lastly, squeeze the Johns Hopkins handle or the Fingertip Release mechanisms on the
rail to disengage the trigger mechanisms while lifting the slide rod(s) up and out of the rail.

Disassembly 

is 

now 

complete. 

SLIDE ROD OR SIDE RAIL/END RAIL REPLACEMENT

1. At the top of one of the Slide Rods, 

fi

 rst lift the portion of the black plastic crib eye cover

that is over the Upper Crib Eye to expose the screw and washer underneath, then loosen
and remove the screw and washer. Repeat at the opposite side of the rail.

Upper Crib Eye

Washer and Screw

Crib Eye Cover

Slide Rod

Bumper Spring

Spacer

Nylon Washers

Slide Rod Adaptor

Lower Crib Eye

14

Содержание 741-CPGP

Страница 1: ...DOERNBECHER CRITICAL CARE CRIB MODELS 741 CPGP 752 KPGP Revised 2022 1 1 31 2022 SERVICE MANUAL The nations 1 manufacturer of hospital cribs 1 800 873 4273 parts hardmfg com...

Страница 2: ...bly Parts List Slide Rod or Side End Rail Replacement Instructions Trigger Assembly for Rails or Johns Hopkins Handle Replacement Trigger Replacement for Portland Dialysis Gates 1904 IV Pole Diagram 1...

Страница 3: ...ard Manufacturing will support parts on our cribs and beds for up to 12 years from the date of manufacture as long as we still make the part or can obtain it from our vendors Cleaning instructions for...

Страница 4: ...l is in place firmly and properly If the trigger mechanism for the side rails fail to lock into place properly remove the crib from service until the mechanism can be replaced SIDE RAIL OPERATION SAFE...

Страница 5: ...position press the touch pad button CRIB UP or DOWN To raise or lower the foot section to desired position press the touch pad button FOOT UP or DOWN To use Trendelenberg or Reverse Trendelenberg spr...

Страница 6: ...the scale off To weigh a patient Follow your hospital protocol to weigh a patient You must ZERO out for an accurate weight reading Remove the child from the crib Press the ZERO button Place the child...

Страница 7: ...ecrib 5 With the 10 kg weight still on the crib select the enter option on the display screen by pressing the Display button If done correctly Cal done should appear on display screen 6 Select the sav...

Страница 8: ...MATTRESS PLATFORM DIAGRAM 2 3 4 7 8 9 Head End 1 Foot End 5 6 10 10 11 10 12 13 14 15 17 18 8...

Страница 9: ...4 Foot Pan Assembly 741 12410 Foot Pan Assembly 752 84159 5 Seat Support Bracket 741 or 752 12544 6 Tilt Indicator 63237 Level for determining angle of elevation 7 Corner Cover 11558 8 Hex Nut 10 32 6...

Страница 10: ...ns Hopkins Handle Assembly 72115 Complete Handle Assembly 6 Screw 10 32x1 RH PH MS 63375 Components at the bottom of the Johns Hopkins Handle 7 Retainer Block 11525 8 Split Lockwasher 63454 9 Acorn Nu...

Страница 11: ...tom trigger Need 2 per side 4 Trigger Assembly Bottom of Portland Gate 84220 1 5 Screw C P 6 1 2 63792 2 For Portland Trigger Need 2 per side Lockwasher not shown 63457 2 6 Portland Spring 63229 2 Com...

Страница 12: ...covering for the tops of the rails Teething Bars 36 wide crib 752 TB 3 Gate only 30 crib head or foot 90791 Gate only 36 crib head or foot 90061 4 Weld assembly 30 head end 82893 U Shaped Frame No Gat...

Страница 13: ...crew 63392 For top of slide rod 5 Nylon Washer 11531 6 Slide Rod Inserts 11567 7 Slide Rod Adaptor 11576 Parts also available in hardware bag 72078 Parts for one slide rod 8 Plastic Washer 1 x 545 x1...

Страница 14: ...ze the Johns Hopkins handle or the Fingertip Release mechanisms on the rail to disengage the trigger mechanisms while lifting the slide rod s up and out of the rail Disassembly is now complete SLIDE R...

Страница 15: ...ins handle or Fingertip Release to lock the rail in place 2 At the bottom of both of the slide rods put the spacer back on first then a washer the bumper spring and the other washer as shown on the pr...

Страница 16: ...trigger bracket inside the rail tube If only the Johns Hopkins Handle assembly is being replaced skip to Step 9 next It is necessary to remove the rail from the crib prior to the steps listed below P...

Страница 17: ...should be in the upright position as shown to the left SIDE TRIGGER ASSEMBLY OR JOHNS HOPKINS HANDLE REPLACEMENT CONTINUED Reassembly is now complete Tab 7 Take the black plastic center trigger brack...

Страница 18: ...rs will be inserted here as well Underside of Gate TRIGGER REPLACEMENT FOR PORTLAND OR DIALYSIS GATES Fingertip Trigger Mechanism Once the broken triggers have been removed from the gate simply insert...

Страница 19: ...1904 IV POLE COMPONENT DIAGRAMS 15 1 2 3 4 5 6 4 5 7 8 9 10 11 12 13 14 Corresponding parts list is on the following page IV Pole Tube 19...

Страница 20: ...asher 2 11555 6 Plastic Spacer 1 62083 7 Release Pin 1 11852 8 Compression Spring 1 63006 9 Plastic Washer 1 11531 Used as release knob 10 Retaining Ring 1 63034 11 Plastic Bracket 2 11553 Available a...

Страница 21: ...l they lock together 5 Align the holes in the bracket and connector then tap in the two connector nuts 14 into the holes in both the top and bottom brackets connectors 6 Insert the two screws 13 into...

Страница 22: ...to Bumper 3 Jam Nylon Locknut 1 4 20 63413 4 4 Washer 1 4 63443 4 5 Bumper Assembly 30 crib 84126 2 Assembly includes 6 7 8 only with bumper plate Bumper Assembly 36 crib 84127 2 6 Grey Corner Bumpers...

Страница 23: ...di erent manner Please contact us for instructions in that situation Suggested Tools 60242 x 6 1 4 20 x 7 8 button socket cap screw 60245 x 6 1 4 20 connector nut 63413 x 4 Jam Nylon Locknut 63443 x 4...

Страница 24: ...the 4 screws 63392 4 locknuts 63413 and washers 63443 located in the 4 corners of the shelf as shown 2 Slide the shelf out of the crib through one of the ends as shown Remove Remove Remove Remove REP...

Страница 25: ...hich crib rests on and are bolted to Assembly is complete REPLACING OR INSTALLING AN UNDER CRIB SHELF OR BUMPER SYSTEM 4 Install the new bumper plate s by aligning the holes in the plate with the hole...

Страница 26: ...4 Bolt 8 8x10mm 61007 2 per caster 5 Casters Full set includes 1 Steer 2 Brake 1 Swivel 3 Caster Sockets BC 40 Not shown This set of casters is on all hospital cribs manufactured prior to 2015 Steer...

Страница 27: ...s of these beds place them in the following categories To avoid overheating the motor after raising and lowering the bed three full times in successions HARD Manufacturing Company requires a fifteen m...

Страница 28: ...vice personnel Caution Based on the results of EMC testing investigations conclude the product tested complies with the requirements of the standard s indicated The results obtained in the test report...

Страница 29: ...ressure washers of any kind Do not allow the crib or bed to air dry as pools of water left on the mattress pan surface or near bolt holes may eventually lead to premature corrosion CRIBS OR BEDS Mattr...

Страница 30: ...1 2020 2 0 Crib Safety Audit 2 1 Filler Bar Spacing Examine the spacing between bars of all cribs in service If a crib has railing spaces that exceed 2 3 8 remove it from service 2 2 Entrapment in cri...

Страница 31: ...c bushings that help the rail glide up and down on the slide rods 3 7 Are the components at the top and bottom of the slide rods intact and free of damage 3 8 Check the sta control and the scale contr...

Страница 32: ...3 0 Physical Inspection of Crib Components Continued YES NO Casters Under Crib Shelf and Bumper System 3 20 Is the crib able to be moved easily do the wheels roll smoothly 3 21 Do the casters show any...

Страница 33: ...anisms squeeze and retract properly 3 31 Are the plastic lift bar glides at either end of each lift bar present and free of damage 3 32 Are the curtain rods at both the top of the curtains and on the...

Страница 34: ...l range of motion 4 4 Press the Hi Lo up down buttons allowing the mattress platform to travel to the top of its range prior to pressing the down button Is it functioning correctly and getting full ra...

Страница 35: ...ould lead to premature corrosion and require replacement of the mattress Expected Service Life of HARD Manufacturing Cribs or Beds is 9 years Terms and Conditions of the Warranty Period The warranty c...

Отзывы: