8. The reset-to-zero calibration lasts for approximately two minutes.
The message
Calibration Step Succeeded
indicates success.
If the display shows
Calibration Step Failed
, check the input value, range, function, and
the input calibration value and repeat the Calibration Steps.
9. Choose the rear input terminal. Install the short circuit block on the rear panel input
terminal.
10. Repeat steps 4 to 9 for the rear input terminal.
11. Switch
DCI, RES and FRES
to complete DC zero calibration.
Note:
Zero calibration for all ranges under the same function is completed in one time,
so there is no need for repeated calibration.
DC voltage gain calibration
Configuration: DC voltage
1. Configure each function and range in the order shown in the table below.
2. Apply the input signal shown in the Input column.
3. Input the input voltage range of the actual application.
The message
Calibration
Step Succeeded
indicates
success.
If the display shows
Calibration Step Failed
,
check the input value, range, function, and the input Calibration value and repeat the
Calibration Step.
4. Repeat steps 1 through 3 for each gain calibration point shown in the table.
The calibration points of HDM3000 are as follows:
gear
Calibration
point 1
Calibration
point 2
100mV
-100mV
100mV
1V
-1V
1V
10V
-10V
10V
100V
-100V
100V
1000V
-1000V
1000V
DC current gain calibration
Configuration: DC current