36
Step 4: For 2-line measurements, the Auto Zero soft key is available.
Auto Zero
:
Auto Zero: Auto Zero provides the most accurate measured value but
requires additional time to perform the reset-to-zero measurement. With Auto
Zero enabled (ON), the DMM makes an internal measurement of the offset after
each measurement and then the offset value is subtracted from the previous
reading.
This prevents the offset voltage on the DMM input circuit from affecting the
measurement accuracy. With Auto Zero disabled (OFF), the DMM makes a
measurement of the offset and subtracts the offset from all subsequent measured
parameters.
Step 5: Press
Aperture
and select the number of power cord circulation (PLC) for
measurement. Only 1, 10 and 100 PLCs provide normal mode (line frequency noise)
suppression.Choose 100 PLC to provide the best noise suppression and resolution,
but the slowest measurement speed:
Step 6: Use the
Units
soft key to display the Celsius, Fahrenheit, or Kelvin
temperature.
The capacitance
This section describes how to configure capacitance measurements from the front
panel.
Step 1: Configure the test leads, as shown below.