DPO6000
,
MPO6000 Series Digital Phosphor Oscilloscope Instruction V1.1
- 5 -
WARNING
This oscilloscope can only be used for measurements within its specified
measurement categories.
Measurement Category Definitions
Measurement category I
is for measurements performed on circuits not directly
connected to MAINS. Examples are measurements on circuits not derived from MAINS,
and specially protected (internal) MAINS derived circuits. In the latter case,transient
stresses are variable. For that reason, the transient withstand capability of the equipment
is made known to the user.
Measurement category II
is for measurements performed on circuits directly connected
to low voltage installations. Examples are measurements on household appliances,
portable tools and similar equipment.
Measurement category III
is for measurements performed in building installations.
Examples are measurements on distribution boards, circuit-breakers, wiring(including
cables, bus-bars, junction boxes, switches and socket-outlets) in fixed installations,
equipment for industrial use and some other equipment. For example,stationary motors
with permanent connection to a fixed installation.
Measurement category IV
is for measurements performed at the source of a low-voltage
installation. Examples are electricity meters and measurements on primary overcurrent
protection devices and ripple control units.
Ventilation Requirement
This oscilloscope uses fan to force cooling. Please make sure that the air intake and
exhaust areas are free from obstructions and have free air. When using the oscilloscope
in a bench-top or rack setting, provide at least 10 cm clearance beside,above and behind
the instrument for adequate ventilation.
WARNING
Inadequate ventilation may cause a temperature increase which can damage
the instrument.So please keep the instrument well ventilated during operation
and inspect the intake and fan regularly.
Working Environment
Temperature
Operating: 0
℃
to +50
℃
Non-operating: -40
℃
to +70
℃
Humidity
0
℃
to +30
℃
: ≤95% relative humidity
+30
℃
to +40
℃
: ≤75% relative humidity
+40
℃
to +50
℃
: ≤45% relative humidity
WARNING
To avoid short circuits inside the instrument or electric shocks, please do not
operate in humid environment.
Altitude
Operating: below 3 km
Non-operating: below 15 km
Care and Cleaning
Care