Temperature High/Low Limit Control:
A user-programmable limit is provided by the Watlow F4 alarms. Alarm 1 is setup for
use with the product thermocouple (input 3). When the limit is tripped the safety valves
and contactors power down to prevent additional temperature changes.
Vibration High Limit Control:
A user-programmable limit is provided by the Watlow F4 alarms. Alarm 2 is setup for
use with the vibration signal (input 2). When the limit is tripped the safety valves power
down and bring the vibration to a halt.
Secondary High Limit Control:
In addition to the user-programmable Hi/Lo limit control there is a Watlow series 97
(Older) or PM EZ-Zone Limit (New). It is a secondary adjustable control is provided to
protect the chamber from high temperature damage. This device senses air
temperature near the electric heat elements in the plenum area and will shut down the
system if plenum temperatures exceed safe limits due to causes such as airflow
obstruction or failure, heat controller failure, or excessive product loading with
insu
ffi
cient cooling (such as occurs when running out of LN2).
This secondary limit can be reset by pressing the reset button located on the controller.
Authorization and Copyright of this document is
controlled by Hanse Environmental Inc.
10
Содержание VTC Series
Страница 1: ...Operation and Maintenance Instructions for Hanse Environmental Inc VTC Series Rev H 2018 ...
Страница 2: ...Blank Page Authorization and Copyright of this document is controlled by Hanse Environmental Inc 2 ...
Страница 34: ...Simulator Circuit Authorization and Copyright of this document is controlled by Hanse Environmental Inc 34 ...
Страница 35: ...Block Diagram Authorization and Copyright of this document is controlled by Hanse Environmental Inc 35 ...
Страница 42: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 42 ...
Страница 56: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 56 ...
Страница 57: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 57 ...
Страница 58: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 58 ...
Страница 59: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 59 ...
Страница 60: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 60 ...
Страница 61: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 61 ...
Страница 62: ...Service Notes Authorization and Copyright of this document is controlled by Hanse Environmental Inc 62 ...
Страница 64: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 64 ...
Страница 65: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 65 ...
Страница 66: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 66 ...
Страница 69: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 69 ...
Страница 70: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 70 ...
Страница 71: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 71 ...
Страница 79: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 79 ...
Страница 83: ...Authorization and Copyright of this document is controlled by Hanse Environmental Inc 83 ...