This error message appears during pump calibration
when the end point can not be reached and the
maximum amount of titrant is exceeded. Check
standard, electrode and/or dosing system and try
again.
The calibration was outside the acceptable limits.
Prepare a new standard and try again.
This error message appears when the pH reading
exceeds the acceptable input limits (-2.00 < pH
< 16.00).
pH CALIBRATION MESSAGES
pH calibration mode.
When the reading has stabilized press
Confirm
to
accept the calibration or
Clear
to restore the default
calibration.
The
“Wrong Buffer”
message is displayed when the
pH value is outside of the acceptable range. Clean the
electrode by following the Cleaning Procedure and/or
check the buffer concentration before continuing the
pH calibration.
Press the
ESC
key to exit pH calibration mode.