27
Note:
• If temperature probe is not connected,
Manual Temperature Compensation
is used
and
MTC
appears on the right side of the screen. If
Automatic Temperature
Compensation
is in use the
ATC
appears on the right side of the screen.
• If an erroneous situation is encountered during the calibration, an error message is
displayed and the calibration can be restarted by pressing
Restart
. Prepare a new
standard, rinse electrode, temperature probe and dosing tip and try again.
• If the calibration doesn’t complete and the max titrant volume of titrant is reached an
error message will be displayed. The calibration can be restarted by pressing
Restart.
Prepare a new standard, rinse electrode, temperature probe and dosing tip and try
again.
• Place the probe holder on the top of the beaker and secure it
by turning clockwise.
• Rinse the pH electrode with deionized water and immerse
into the sample until the Teflon reference junction is completely
submerged. Be sure that the tip of the electrode is not hitting
the stir bar. If necessary additional distilled or deionized
water can be added.
• Insert the dosing tip into the titrant tube sleeve. IT IS CRITICAL
THAT THE TIP BE IMMERSED APPROXIMATELY 0.5 CM (0.1")
INTO THE SOLUTION BEING TITRATED.
• Press
Continue
to begin the calibration and
Stop
to abort it.
• At the end of the calibration, ”
Calibration Completed
“
appears on display. To repeat the calibration press
Restart
or
ESC
to return to the main screen.