OPERATIONAL GUIDE

ORP electrode CONNECTION

The BNC connector is protected
against vapors and humidity by a wa-
terproof sheath.
• Slide the protective sheath down.

Connect the HI  3214P/2 ORP elec-
trode to the BNC connector and
then slide up. Make sure the con-
nector is completely covered.

• Do not be alarmed if white crystals appear around the

electrode protective cap. This is normal with ORP elec-
trodes and they dissolve when rinsed with water.

HI 982401 ORP 

Range

±999 mV

Resolution

1 mV

Accuracy (@20°C/68°F)

±5 mV

Typical EMC Deviation

±5 mV

Setpoint

Selectable from 350 to 850 mV

Alarm

LED blinks when the measured

ORP value is lower than the setpoint

Calibration

Factory calibrated

Probe

HI  3214P/2 interchangeable ORP electrode

(optional)

Casing

IP54

Power Supply

12 Vdc adapter (included)

Dimensions

86 x 110 x 43 mm (3.4 x 4.3 x 1.7”)

Weight

150 g (5.3 oz.)

SPECIFICATIONS

3

P

R

• When not in use, rinse the electrode with water and store

it with the protective cap.

• If the electrode has been left dry, soak the tip in HI 70300

storage solution for at least one hour to reactivate it.

Taking ORP Measurements

• Turn the meter on.
• Remove the protective cap from the

ORP electrode and immerse the tip
(4 cm/1½”) in the sample.

• The LCD will show the ORP value

measured in mV unit. Allow the read-
ing to stabilize and HI 982401 will
start continuous monitoring.

Note: to prevent damages to the meter,

remove the ORP electrode from
the sample before turning the
meter off.

Adjusting the Setpoint

• To access the MEASURE/SET switch

and the setpoint adjustment trimmer,
unscrew the 4 screws on the meter
back and remove the rear panel.

• Move the switch to the SET position

(left).

4

Содержание Pronto HI 982401

Страница 1: ... accident misuse tampering or lack of prescribed maintenance are not covered If service is required contact the dealer from whom you purchased the instrument If under warranty report the model number date of purchase serial number and the nature of the failure If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Inst...

Страница 2: ... longer life even in harsh conditions GENERAL DESCRIPTION 1 HI 3214P 2 Double fiber junction plastic body Platinum ORP electrode with 2 m 6 6 cable and BNC connector HI 7020M ORP test solution at 200 275 mV 230 mL HI 7020L ORP test solution at 200 275 mV 500 mL HI 7091M Reducing solution 230 mL HI 7091L Reducing solution 500 mL HI 7092M Oxidizing solution 230 mL HI 7092L Oxidizing solution 500 mL ...

Страница 3: ...er while mak ing sure the gasket is in place before tightening the screws Whenever the mV reading is lower than the setpoint the red alarm LED blinks to warn the user Maintenance To minimize clogging and provide longer life to the ORP electrode it si recommended to clean it monthly Immerse the tip of the electrode in HI 7061 cleaning solution for at least one hour and then rinse it with tap water ...

Страница 4: ...trode optional Casing IP54 Power Supply 12 Vdc adapter included Dimensions 86 x 110 x 43 mm 3 4 x 4 3 x 1 7 Weight 150 g 5 3 oz SPECIFICATIONS 3 P R When not in use rinse the electrode with water and store it with the protective cap If the electrode has been left dry soak the tip in HI70300 storage solution for at least one hour to reactivate it Taking ORP Measurements Turn the meter on Remove the...

Отзывы: