46
PC CONNECTION
• status (1 char): N (new) - calibrated in last calibration; O (old) - from an
old calibration.
• warnings during calibration (2 chars): 00 - no warning.
• standard value, with sign and decimal point and exponent (11 chars).
• calibration time,
yymmddhhmmss
(12 chars).
PAR
Requests the setup parameters setting.
The answer string contains:
• Instrument ID (4 chars)
• Calibration Alarm time out for pH (2 chars)
• Calibration Alarm timeout for ISE (2 chars) - if ISE available
• SETUP information (2 chars): 8 bit hexadecimal encoding.
• 0x01 - beep ON (else OFF)
• 0x04 - degrees Celsius (else degrees Fahrenheit)
• 0x08 - Offset calibration (else Point calibration)
• Auto Light Off time (3 chars)
• Auto Power Off time (3 chars)
• The number of custom buffers (1 char)
• The custom buffer values, with sign and decimal point, for each defined custom
buffer (7 chars)
• The ID of the ISE electrode (2 chars) - if ISE available
• The molar weight of the selected ION, with sign and decimal point (9 ASCII
characters)
• The ion charge (2 chars)
• The ISE unit (2 chars)
• The short name of the selected language (3 chars)
NSLx
Requests the number of logged samples (4 chars).
The command parameter (1 char):
• P - request for pH range
• M - request for mV and Rel mV ranges
• I - request for ISE range
LODPxxx
Requests the xxxth pH record logged data.
LODMxxx
Requests the xxxth mV/Rel mV record logged data.
LODIxxx
Requests the xxxth ISE record logged data.
LODPALL
Requests all pH Log on demand.
LODMALL
Requests all mV/Rel mV Log on demand.