29
CALIBRA
TION
Press
Change
to enter buffer edit mode.
Press / keys to change the buffer value.
Press
Accept
to confirm new value or
ESC
to exit edit mode.
Clear Calibration
Press
Clear
functional key to clear previous calibrations. Calibration points saved during current
calibration are not deleted.
Note:
If Clear calibration is invoked during the first calibration point, the instrument returns to
measurement mode.
Electrode Condition
If enabled (see METER SETUP section), the display will show the probe icon and electrode condition
(%) after a successful calibration. The electrode condition is evaluated only if current calibration
includes at least two standard buffers and remains active until the end of the calibration day.