15
14
Code
Valid Values
Default
PW
92 EEPROM selftest
0: off
0
yes
1: on
93 Relays and LEDs selftest
0: off
0
yes
1: on
94 Watchdog selftest
0: off
0
yes
1: on
99 Unlock password
0000 to 9999
0000
yes
Note
The process controller automatically checks to ensure that
the entered data matches other related variables. If a wrong
configuration is entered, “WRONG” blinks on the LCD to
prompt the user. The correct configurations are the following:
If M1=
/ 0 then S1<HA, S1>LA;
If M2=
/ 0 then S2<HA, S2>LA;
If M1= 1 then S1-H1>LA;
If M1= 2 then S1+H1<HA;
If M1= 3 then S1+D1<HA;
If M1= 4 then S1-D1>LA;
If M2= 1 then S2-H2>LA;
If M2= 2 then S2+H2<HA;
If M2= 3 then S2+D2<HA;
If M2= 4 then S2-D2>LA;
If M1= 1 and M2 = 2
then S1-H1>S2+H2, S2>LA, HA>S1;
If M1 = 2 and M2 = 1
then S2-H2>S1+H1, S1>LA, HA>S2;
If M1 = 3 and M2 = 2
then S1>S2+H2, S2>LA, HA>S1+D1;
If M1 = 2 and M2 = 3
then S1+H1<S2, S1>LA, HA>S2+D2;
If M1 = 4 and M2 = 1
then S1<S2–H2, S1–D1>LA, HA>S2;
If M1 = 1 and M2 = 4
then S1–H1>S2, S2–D2>LA, HA>S1;
If M1 = 3 and M2 = 4
Code
Valid Values
Default
PW
30 Relay 3 high alarm (HA) 0.00 to 14.00 pH
9.00 pH
no
-2000 to 2000 mV
600 mV
HA>LA, HA>S1 or HA>S2
31 Relay 3 low alarm (LA)
0.00 to 14.00 pH
5.00 pH
no
-2000 to 2000 mV
-600 mV
LA<HA, LA<S1 or LA<S2
32 Proportional control
1 to 30 min
5
no
mode period
33 Maximum relay ON time 10 to 9999 min
60
no
(after which an alarm mode is entered)
40 Analog output selection
0: 0-1mA
2
no
1: 0-20 mA
2: 4-20 mA
3: 0-5 VDC
4: 1-5 VDC
5: 0-10 VDC
41 Analog output
0.00 to 13.00 pH
0.00 pH
no
lower limit
-2000 to 2000 mV
-2000 mV
(O_VARMIN)
(O_VARMIN< O_VARMAX - (1.00pH or 50mV))
42 Analog output
1.00 to 14.00 pH
14.00 pH
no
upper limit
-2000 to 2000 mV
2000 mV
(O_VARMAX)
(O_VARMIN< O_VARMAX - (1.00pH or 50mV))
60 Current day
01 to 31
from RTC
no
61 Current month
01 to 12
from RTC
no
62 Current year
1997 to 9999
from RTC
no
63 Current time
00:00 to 23:59
from RTC
no
71 Baud rate
1200, 2400, 4800, 9600
4800
no
90 Display selftest
0: off
0
yes
1: on
91 Keyboard selftest
0: off
0
yes
1: on