• Press or again to view the second memorized buffer
at the time of last calibration. The secondary display will
show "BUF2" to indicate second buffer.
• Press or again to view the third memorized buffer at
the time of last calibration. The secondary display will
show "BUF3" to indicate third buffer.
• Press or again to return to the first CAL DATA display
(date) at the time of last calibration.
Note
In any moment, by pressing LCD or CAL DATA the meter will
return to the regular operating display.
• Press or to view the time of last calibration. The sec-
ondary display will show "HOU" to indicate hours.
• Press or again to view the offset in mV at the time of
last calibration. The secondary display will show "OFF"
to indicate offset.
• Press or again to view the first slope in mV at the time
of last calibration. The secondary display will show "SL1" to
indicate first slope.
• Press or again to view the second slope in mV at the
time of last calibration. The secondary display will show
"SL2" to indicate second slope.
• Press or again to view the first memorized buffer at
the time of last calibration. The secondary display will
show "BUF1" to indicate first buffer.