background image

13

12

For refillable electrodes

If the filling solution (electrolyte) is more than 2.5 cm (1") below the
fill hole, add 

HI 7082 

3.5M KCl electrolyte solution for double junction

or 

HI 7071 

3.5M KCl+AgCl electrolyte solution for single junction

electrodes.
For a faster response, unscrew the fill hole screw during measure-
ments.

For AmpHel

®

   electrodes

If the electrode does not respond to pH changes, the battery is run
down and the electrode should be replaced.

MEASUREMENT

Rinse the electrode tip with distilled water. Immerse the tip (bottom 4
cm /1½") in the sample and stir gently for a few seconds.
For a faster response and to avoid cross contamination of the samples,
rinse the electrode tip with a few drops of the solution to be tested,
before taking measurements.

STORAGE

To minimize clogging and assure a quick response time, the glass bulb
and the junction should be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of

HI 70300 

storage solution or, in its absence, filling solution (

HI 7071

for single junction or 

HI 7082 

for double junction electrodes). Follow

the Preparation Procedure above before taking measurements.

Note:

NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.

PERIODIC MAINTENANCE

Inspect the electrode and the cable. The cable used for connection to
the meter must be intact and there must be no points of broken
insulation on the cable or cracks on the electrode stem or bulb.
Connectors must be perfectly clean and dry. If any scratches or cracks
are present, replace the electrode. Rinse off any salt deposits with
water.

For refillable electrodes:

Refill the reference chamber with fresh electrolyte (

HI 7071 

for single

junction or 

HI 7082 

for double junction electrodes). Allow the elec-

trode to stand upright for 1 hour.
Follow the Storage Procedure above.

AmpHel

®

 is a registered Trademark of "Hanna Instruments"

ELECTRODE CONDITIONING & MAINTENANCE

PREPARATION

Remove the protective cap.
DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT. This is normal
with electrodes and they will disappear when rinsed with water.
During transport tiny bubbles of air may form inside the glass bulb
affecting proper functioning of the electrode. These bubbles can be
removed by "shaking down" the electrode as you would do with a
glass thermometer.
If the bulb and/or junction is dry, soak the electrode in 

HI 70300

storage solution for at least one hour.

Note:

 To prevent damage to the pH electrode, remove it from the

solution before turning the meter off.

With the meter OFF, detach the electrode from the instrument
and immerse it in the storage solution.

Содержание pH 211R

Страница 1: ...Instruction Manual pH211R Microprocessor based Bench pH mV C Meter with Analog Output w w w h a n n a i n s t c o m ...

Страница 2: ...emperature values simultaneously The user is guided through the calibration process with simple instructions on the display The meter also features 0 to 5 V analog output TABLE OF CONTENTS PRELIMINARY EXAMINATION 3 GENERAL DESCRIPTION 3 FUNCTIONAL DESCRIPTION 4 SPECIFICATIONS 5 OPERATIONAL GUIDE 6 pH CALIBRATION 9 ELECTRODE CONDITIONING MAINTENANCE 12 TEMPERATURE RESISTANCE CORRELATION FOR pH SENS...

Страница 3: ...utput connector 16 Negative analog output connector SPECIFICATIONS Range 2 00 to 16 00 pH 399 9 2000 mV 9 9 to 120 0 C Resolution 0 01 pH 0 1 1 mV 0 1 C Accuracy 20 C 68 F 0 01 pH 0 2 1 mV 0 5 C Typical EMC Deviation 0 03 pH 2 mV 0 3 C pH Calibration Automatic 1 or 2 point with 5 memorized buffers pH 4 01 6 86 7 01 9 18 10 01 Temperature Compensation Automatic with HI 7669 2W probe or manual 9 9 t...

Страница 4: ... the oxidizing or reducing power of the sample tested To correctly perform a redox measurement the surface of the ORP electrode must be clean and smooth OPERATIONAL GUIDE Power Connection Plug the 12 Vdc adapter into the power supply socket Note Make sure the mains line is protected by a fuse Electrode and Probe Connections Connect the BNC of the pH or ORP electrode to the BNC socket on the rear p...

Страница 5: ...2 5 V 5 V RESET The RESET button 14 on page 4 should only be used when the instrument displays erroneous messages due to strong electrical inter ference or when the power supply was disconnected before switching the meter off After pressing RESET always recelibrate the unit before proceeding pH CALIBRATION Frequent calibration is recommended Moreover the instrument should be re calibrated for pH a...

Страница 6: ... the electrode by following the maintenance procedure see Elec trode Maintenance section If necessary change buffer or electrode Note Press RANGE to display the temperature reading on the LCD during calibration One Point Calibration Immerse pH electrode and temperature probe approximately 4 cm 1 into any of the available buffer solutions pH 4 01 6 86 7 01 9 18 or 10 01 and stir gently The temperat...

Страница 7: ...DISTILLED OR DEIONIZED WATER PERIODIC MAINTENANCE Inspect the electrode and the cable The cable used for connection to the meter must be intact and there must be no points of broken insulation on the cable or cracks on the electrode stem or bulb Connectors must be perfectly clean and dry If any scratches or cracks are present replace the electrode Rinse off any salt deposits with water For refilla...

Страница 8: ...4 inorganic cleaning solution for 15 minutes Oil grease Rinse with HI 7077 Oil Fat cleaning solution IMPORTANT After performing any of the cleaning procedures rinse the electrode thoroughly with distilled water refill the reference cham ber with fresh electrolyte not necessary for gel filled electrodes and soak the electrode in HI 70300 storage solution for at least 1 hour before taking measuremen...

Страница 9: ...bottle 4 x 50 mL HI 7082 3 5M KCl electrolyte solution 4 x 50 mL bottle for double junction electrodes HI 8082 3 5M KCl electrolyte solution in FDA bottle 4 x 50 mL for double junction electrodes TROUBLESHOOTING GUIDE Symptoms Problem Solution The meter is slow in responding or gives faulty readings The meter does not ac cept the 2nd buffer solution for calibration The reading drifts Display shows...

Страница 10: ...m DIA 0 29 S VERSION HI 1331 ORP PRETREATMENT SOLUTIONS HI 7091L Reducing pretreatment solution 500 mL bottle HI 7092L Oxidizing pretreatment solution 500 mL bottle pH ELECTRODES All electrode part numbers ending in B are supplied with BNC connector and 1 m 3 3 cable HI 1043B Glass body double junction refillable combination pH electrode Use strong acid alkali HI 1053B Glass body triple ceramic co...

Страница 11: ... 5 HI 4430 S VERSION FC 100B PVDF body double junction refillable combination pH electrode Use general purpose for food industry FC 200B PVDF body open junction conic Viscolene non refillable combination pH electrode Use meat cheese FC 210B Glass body double junction conic Viscolene non refillable combina tion pH electrode Use milk yogurt FC 220B Glass body triple ceramic single junction refillabl...

Страница 12: ...able 3 m 9 9 long OTHER ACCESSORIES HI 710005 Voltage adapter 115 Vac 12 Vdc US plug HI 710006 Voltage adapter 230 Vac 12 Vdc European plug HI 710012 Voltage adapter 230 Vac 12 Vdc UK plug HI 710013 Voltage adapter 230 Vac 12 Vdc S Africa plug HI 710014 Voltage adapter 230 Vac 12 Vdc Australian plug HI 76404 Electrode holder HI 7669 2W Temperature probe with 1 m 3 3 cable HI 8427 pH ORP electrode ...

Страница 13: ...941 Fax 21 4584 2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2983 Thailand Tel 66 2619 0708 Fax 66 2619 0061 UnitedKingdom Tel 01525 850 855 Fax 01525 853 668 USA Tel 401 765 7500 Fax 401 7...

Отзывы: