To adjust the light level of the blank sample:
1. Press DIAG. The display will show the last diagnostic mode
and the right-most numerical digit will blink.
2. Enter code 999
using the
and
keys. After the code is
selected display will show “d999”.
3. Press DIAG to execute the light level set mode. All the
display digits will blink.
4. The display will show direct reading of the photodetector
cell.
5. Make sure the photometric chamber is well flushed. Stop
the external sample stream from flowing into the system.
The photometric cell is now filled with blank sample solu-
tion.
6. Adjust the trimmer on the power supply board to obtain
a reading between 7300 and
7700.
7. Press any key to leave this mode.