17
Grasp the other plastic collar and pull, stretching the center
section, and place the grommet in the lower left indentation
hole.
Repeat this process with the second pump tube, placing it in
the upper indentation holes.
Separate reagent caps are provided in the accessory kit. Put
the supplied caps onto each reagent bottle prior to installing
them. Place the indicator bottle (
HI 70450
for
PCA 300
and
HI 70460
for
PCA 301
) on the right and the buffer
bottle (
HI 70451
for
PCA 300
and
HI 70461
for
PCA
301
) on the left.
Note
Add the content of 5
HI 70452
sachets, DPD Compound,
to the Indicator Solution prior to installing it.
Connect the longer tube
ends on the left side of
the pump to the reagent
bottle cap insert fitting.
Connect the short ends
on the right side of the
pump to the measuring
cell reagent input port
through the "Y" connec-
tor.