38
• Fill the cuvet up to the mark with 10 mL of
sample and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and "SIP" will appear on the display.
• Wait for a few seconds and the display will
show "0". Now the meter is zeroed and ready
for measurement.
• Fill a beaker up to the
25 mL mark with the
sample, add the con-
tent of one packet of
HI 93722 reagent and
stir gently to mix.
• Fill a second cuvet with 10 mL of the reacted
sample up to the mark. Replace the cap.
• Reinsert the cuvet into the instrument.
• Wait for 45 seconds and press READ. "SIP" will appear during
measurement.
• The instrument directly displays concentration in mg/L of cyanuric
acid on the Liquid Crystal Display.