29
C 104 - PH, CL
C 104 - PH, CL
C 104 - PH, CL
C 104 - PH, CL
C 104 - PH, CL
22222
AND CYANURIC ACID
AND CYANURIC ACID
AND CYANURIC ACID
AND CYANURIC ACID
AND CYANURIC ACID
C 104
pH/Cl/Cyanuric A.
The
C 104
meter measures pH, chlorine (Cl
2
, free and total) and
cyanuric acid content in water and wastewater.
SPECIFICATIONS
Range
pH 5.9 to 8.5
Free Cl
2
0.00 to 2.50 mg/L
Total Cl
2
0.00 to 3.50 mg/L
Cyanuric Acid 0 to 80 mg/L
Resolution
pH 0.1
Cl
2
0.01 mg/L
Cyanuric Acid 1 mg/L
Accuracy
pH ±0.1
Cl
2
±0.03 mg/L ±3% of reading
Cyanuric Acid ±1 mg/L ±15% of reading
Typical EMC
pH ±0.1
Deviation
Cl
2
±0.01 mg/L
Cyanuric Acid ±1 mg/L
Light Source
Light Emitting Diode @ 555 nm
Method
For chlorine, adaptation of the EPA recom-
mended DPD method 330.5. For pH, Phenol
red method. For cyanuric acid, adaptation of
the turbidimetric method.
REQUIRED REAGENTS
Code
Unit
Description
Quantity
HI 93710-0
pH
Phenol red
5 drops
HI 93701-0
Free Cl
2
DPD
1 packet