21
Autodelete BOD start data
Highlight
Autodelete BOD start data.
Press the displayed functional key to enable/disable the feature.
If enabled the
BOD
initial data record used in
BOD
result evaluation is deleted automatically after the
BOD
result has been saved into the meter’s memory (
LOG
key pressed).
If disabled, the user has to delete
BOD
initial data records that were used in
BOD
result evaluation,
entering
View initial BOD data
mode
.
Press / keys to modify the selected parameter’s value.
Press
Prev/Next
to select a different parameter.
Press
Save
to save the new
SOUR
configuration.
Press
ESC
to leave without changing.
•
Min time
‑ the minimum time for the
SOUR
test.
Range: 1 to 3600 seconds.
•
Max time
‑ the maximum time for the
SOUR
test. The test will stop automatically when the
maximum time has elapsed.
Range: 1 to 3600 seconds.
•
Min start
DO
‑ the minimum accepted
DO
value for starting the
SOUR
test. If the
DO
reading is
less than this value the test cannot be started.
Range: 0.01 to 50.00 mg/L.
•
Min end DO
‑ the minimum accepted
DO
value at the end of the test. If the
DO
reading at the
end of the
SOUR
test is less than this value a warning message will be displayed.
Range: 0.00 to 50.00 mg/L.
•
Total volume
‑ the volume of the diluted mixture.
Range: 0.1 to 300.0 mL
•
Sample volume
‑ the volume of sample in the diluted mixture.
Range: 0.1 to 300.0 mL.
•
Solids weight:
Total solids or Volatile suspended solids weight.
Range: 0.1 to 300.0 g/L.
•
SOUR @ 20°C:
If this option to enabled the
SOUR
value is corrected to 20 °C.