54
• Press
CFM
to confirm.
• The second expected calibrated point is displayed.
• Immerse the probe into the second vessel as close as possible to the reference thermometer. Allow
a few seconds for the probe to stabilize.
• Use the
/
keys to set the calibration point value to that measured by the reference
thermometer.
• When the reading is stable and within range of the selected calibration point,
CFM
functional key
is displayed.
• Press
CFM
to confirm. The instrument returns to measurement mode.
TEMPERA
TURE CALIBRA
TION (for technical personnel only)