61
METHOD PROCEDURES
61
61
OZONE
10.12. OZONE
SPECIFICATIONS
Range
0.00 to 2.00 mg/L (as O
3
)
Resolution
0.01 mg/L
Accuracy
±0.02 mg/L ±3% of reading at 25 °C
Light Source
LED with narrow band interference filter @ 525 nm
Method
Colorimetric DPD Method
REQUIRED REAGENTS
Code Description
Quantity
HI93757-0
Ozone Reagent
1 packet
HI93703-52-0
Glycine Powder (Optional Reagent) 1 packet
REAGENT SETS
HI93757-01
Reagents for 100 tests
HI93757-03
Reagents for 300 tests
HI93703-52
Reagents for 100 tests (Optional)
For other accessories see ACCESSORIES section.
STANDARD MEASUREMENT PROCEDURE
Chlorine free samples
• Select the
Ozone
method using the procedure described in the METHOD SELECTION section.
• Fill the cuvette with 10 mL of unreacted sample (up to the mark).
Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and close the lid.
• Press
Zero
. The display will show “-0.0-” the meter is zeroed and ready for measurement.
10 mL