45
METHOD PROCEDURES
45
45
CHLORINE
, T
OT
AL
10.6. CHLORINE, TOTAL
SPECIFICATIONS
Range
0.00 to 5.00 mg/L (as Cl
2
)
Resolution
0.01 mg/L
Accuracy
±0.03 mg/L ±3% of reading at 25 °C
Light Source
LED with narrow band interference filter @ 525 nm
Method
Adaptation of the EPA DPD Method 330.5
REQUIRED REAGENTS
POWDER
Code Description
Quantity
HI93711-0
Total Chlorine Reagent
1 packet
LIQUID
Code Description
Quantity
HI93701A-T
Total Chlorine Reagent A
3 drops
HI93701B-T
Total Chlorine Reagent B
3 drops
HI93701C-T
Total Chlorine Reagent C
1 drop
REAGENT SETS
HI93701-T
Reagents for 300 tests (liquid)
HI93711-01
Reagents for 100 total tests (powder)
HI93711-03
Reagents for 300 total tests (powder)
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
• Select the
Chlorine (Total)
method using the procedure described in the METHOD SELECTION
section.
• Fill the cuvette with 10 mL of unreacted sample (up to the mark).
Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and close the lid.
10 mL