13

12

When the battery level is below 10%, a warning
symbol will blink on the LCD to indicate a low
battery condition.
If the battery level is low enough to cause erroneous readings, the
Battery Error Prevention System (BEPS) turns the meter off.
Immediately replace the batteries with new ones.
The batteries are accessed by separating the front and the back
halves of the meter: unscrew the 4 screws on the back of the meter
and carefully replace the three batteries located in the battery
compartment, while paying attention to their polarity. Reattach the
back making sure that the gasket is in place and tighten the 4
screws to ensure a watertight seal.
Battery replacement must only take place in a non-hazardous area
using 1.5V AA (IEC LR6) alkaline batteries.

BATTERY REPLACEMENT

USER CALIBRATION

The 

HI 93510N

 and 

HI 93522

 models can be calibrated at 0°C by

using an ice bath.
• Prepare an ice bath with approximately equal volumes of distilled

water and chopped ice made from distilled water.

• Immerse the temperature probe in the center of the ice bath,

taking care not to touch the ice with the probe tip.

• Ensure that the meter is measuring a temperature within ±3°C.
• To enter the Calibration mode, press and hold the CAL button for

about 5 seconds.

• The CAL tag turns on to indicate that

the Calibration mode has been entered.

Note

: If the measurement is outside the

±3°C window, the meter does not
enter Calibration mode.

• When the meter reaches the stability condition, which is detected

when the measurement remains within ±0.2°C for 5 seconds, the
calibration is accepted and the reading becomes 0°C (32°F).

• The meter then automatically returns to normal mode.

Note

: To exit the Calibration mode at any time, press the CAL key.

Note

: User calibration cannot be entered in relative or in T1-T2 mode.

Note

: User calibration is only performed on the current channel

displayed (T1 or T2).

All Hanna thermometers have been accurately pre-calibrated at the
factory. It is generally recommended to have all thermometers recali-
brated at least once a year. For an accurate recalibration, contact your
nearest Hanna Customer Service Center.

Hanna Accuracy Test Plugs

 provide a quick and easy way to check the

meter's accuracy by simply connecting these Test Plugs to the probe
connector of the meter. If the reading differs by more than ±0.4°C
(±0.8°F) from the Test Plug value, the meter is due for recalibration.

Choose the right Test Plug to suit your needs:

HI 762-18C

Calibration test key, -18.0°C

HI 762000C

Calibration test key, 0.0°C

HI 762070C

Calibration test key, 70.0°C

HI 762-004F

Calibration test key, -0.4°F

HI 762032F

Calibration test key, 32.0°F

HI 762158F

Calibration test key, 158.0°F

FACTORY RECALIBRATION

Recommendations for Users

Before using these products, make sure that they are entirely suitable for the
environment in which they are used.
Operation of these instruments in residential area could cause unacceptable inter-
ferences to radio and TV equipments, requiring the operator to take all necessary
steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments' EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the mea-
surement surface exceed 24VAC or 60 VDC.
To avoid damages or burns, do not perform any measurement in microwave ovens.

Note: To clean the meters, do not use aggressive detergents. It is

recommended to use water.

Содержание HI 93510N

Страница 1: ...Instruction Manual HI 93510 HI 93510N HI 93512 HI 93522 PortableMicroprocessor ThermistorThermometers w w w h a n n a i n s t c o m...

Страница 2: ...and probe can be calibrated in an ice bath at 0 C or dimly lit areas the backlight feature can be easily activated through the keyboard HI 93512 is a two channel thermometer ideal for monitoring two...

Страница 3: ...obes series Available HI 762 accuracy test plugs Display of temperature in degrees Celsius or Fahrenheit Range 50 to 150 C or 58 to 302 F HOLD function HI and LO Max and Min temperature values always...

Страница 4: ...with HI 762BL probe Battery 3x1 5V AA IEC LR6 batteries approx 2000 hours of continuous use BL off Auto off user selectable 8 min 60 min disabled HI 93522 only Environment 10 to 50 C 14 to 122 F RH 1...

Страница 5: ...mpartment see figure on page 10 Unscrew the 4 screws on the back of the meter and select the desired auto off time period by pressing the button and viewing the value on the LCD Reattach the back maki...

Страница 6: ...light up to inform the user To select the desired channel use the T1 T2 key The 2 channel models are also provided with the DISP function which allows the user to select the information to be displaye...

Страница 7: ...then automatically returns to normal mode Note To exit the Calibration mode at any time press the CAL key Note User calibration cannot be entered in relative or in T1 T2 mode Note User calibration is...

Страница 8: ...obe w 10 m 33 cable white handle HI 762PBL General purpose penetration probe w 1 m 3 3 cable blue handle HI 762PBL 10 General purpose penetration probe w 10 m 33 cable blue handle HI 762PG General pur...

Страница 9: ...42 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 0...

Отзывы: