TIPS FOR ACCURATE OPERATION

The instructions listed below should be carefully followed during testing and during
calibration.
• All glassware that comes into contact with standards should be maintained

clean. Wash with 

HI93703-50

 cleaning solution and rinse with 

HI 93124-0

or turbidity free water.

• For consistent readings, it is recommended to perform measurement immediately

after collecting sample.

• To obtain a representative sample, gently mix the liquid before samples are

taken. Do not shake (to prevent air bubbles) and do not let the sample settle.

• Rinse the cuvette twice with 5 ml of the liquid to be tested for removing the effect

of dust or any previous liquid. Gently pour the liquid down the side of the cuvette
to reduce air bubbles.

• Try to tighten the cuvette cap always to the same degree.

• Before inserting the cuvette into the instrument, wipe it with soft, lint-free tissue.

Handle cuvettes so that no fingerprints can get on the areas where light passes
(approximately 2 cm/1 inch from the bottom of the cuvette).

• Discard the sample soon after the reading is taken to avoid permanently clouding

the glass.

If any problem is experienced during measurements, please contact your dealer or
the nearest Hanna Instruments Customer Service Department.

LCD ERROR CODES

Every time the meter is switched on, the RTC (Real Time Clock) and the internal
memory (EEPROM) are tested.

If an error occurs, the corresponding error code will be displayed.

Here below are listed the error codes:

No cover error; check cuvette position

Calibration error; check  calibration standard
solution

Real Time Clock  error (*)

EEPROM error (*)

Internal communication error (*)

Internal Bus error (*)

(*) Contact the nearest Hanna Instruments Customer Service Department.

Содержание HI 93124

Страница 1: ...1 Instruction Manual HI 93124 Portable Logging Haze Meter w w w h a n n a i n s t c o m...

Страница 2: ...instructions This warranty is limited to repair or replacement free of charge Damage due to accidents misuse tampering or lack of prescribed maintenance are not covered If service is required contact...

Страница 3: ...epolyphenol polymerizes A range of stabilization treatments is available for avoiding haze problems and theproducthastobecontrolledonseveralstepsduringbrewingprocess inparticular after filtration and...

Страница 4: ...withalint freetissue HI731318 before inserting into the measure ment cell Thecuvettemustbecompletelyfreeoffingerprints andotheroilordirt particularlyintheareawhere the light goes through approximately...

Страница 5: ...availableinstrumentsoperating in the range of visible light Air bubbles and the effect of vibrations that disturb the surface of the sample willgivefalseresults Dirtyglasswarecouldalsoaffectreadingsal...

Страница 6: ...from stock solutions TheconsistencyofHI93124readingsbyusingbothstandardshasbeenseparately established by Advanced Polymer Systems and Hanna Instruments Additional documentation about the formazine sta...

Страница 7: ...ette cap always to the same degree Beforeinsertingthecuvetteintotheinstrument wipeitwithsoft lint freetissue Handlecuvettessothatnofingerprintscangetontheareaswherelightpasses approximately 2 cm 1 inc...

Страница 8: ...4 1 HI 93124 2 HI731318 Tissue for wiping cuvettes 4 pcs HI 731321 Spare measurement cuvettes 4 pcs HI92000 Windows compatibleapplicationsoftware HI920011 5 to 9 pin RS232 connection cable HI 93703 50...

Страница 9: ...3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiw...

Отзывы: