28
•
Immerse the conductivity probe into the
50°C (122°F) bath and wait for a couple
of minutes for thermal equilibrium.
•
Press CAL to display the "50.0°C", "CAL" symbol and an intermit-
tent "BUF" indicator. "
" will symbol blink for a few seconds.
•
"
" will disappear from the display
when the reading is stable and "CON" will
start blinking.
•
Press CFM to confirm the calibration value.
The calibration is now complete and the instrument returns to normal
operation.
Note:
You can quit the calibration mode at any
time by pressing ON/OFF.
°C
50 °C
(122 °F)
CAL
BUF
CAL
°C
°C
CON
CFM
ON/OFF