23
°C
°F
HI 7032
HI 7036
mg/L
g/L
(ppm)
(ppt)
0
32
758
6.82
5
41
876
7.88
10
50
999
8.99
15
59
1122
10.10
16
60.8
1148
10.33
17
62.6
1173
10.56
18
64.4
1200
10.78
19
66.2
1224
11.01
20
68
1251
11.24
21
69.8
1277
11.47
22
71.6
1303
11.71
23
73.4
1329
11.94
24
75.2
1358
12.18
25
77
1382
12.41
26
78.8
1408
12.65
27
80.6
1438
12.89
28
82.4
1461
13.13
29
84.2
1476
13.37
30
86
1515
13.61
31
87.8
1541
13.85
The TDS value in aqueous solutions is directly proportional to conduc-
tivity. The ratio between the two parameters depends on the solution
and usually it is set to a factor of 0.5 (corresponding to a solution of
CaCO
3
). This means that 1 µS/cm is equal to 0.5 mg/L (ppm) of TDS.
With
HI 9635
,
the TDS factor can be set to 0.5 or 0.7 depending on
the solution to be measured (see page 26). For accurate TDS
measurements, make sure that
HI 9635
has been calibrated in a
conductivity standard solution and the temperature coefficient is
properly set.
TDS VERSUS
TDS VERSUS
TDS VERSUS
TDS VERSUS
TDS VERSUS
TEMPERATURE CHART
TEMPERATURE CHART
TEMPERATURE CHART
TEMPERATURE CHART
TEMPERATURE CHART