A microprocessor controlled special tungsten lamp emits radiation which is first optically conditioned
and beamed to the sample contained in the cuvet. The optical path is fixed by the diameter of the
cuvet. Then the light is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of
intensity 

I

o

 or 

I.

The photoelectric cell collects the radiation 

I

 that is not absorbed by the sample and converts it

into an electric current, producing a potential in the mV range.
The microprocessor uses this potential to convert the incoming value into the desired measuring
unit and to display it on the LCD.
The measurement process is carried out in two phases: first the meter is zeroed and then the actual
measurement is performed.
The cuvet has a very important role because it is an optical element and thus requires particular
attention. It is important that both the measurement and the calibration (zeroing) cuvets are
optically identical to provide the same measurement conditions. Whenever possible use the same
cuvet for both. It is necessary that the surface of the cuvet is clean and not scratched. This to avoid
measurement interference due to unwanted reflection and absorption of light. It is recommended
not to touch the cuvet walls with hands.
Furthermore, in order to maintain the same conditions during the zeroing and the measuring
phases, it is necessary to close the cuvet to prevent any contamination.

US Environmental Protection Agency
degree Celsius
degree Fahrenheit
milligrams per liter. mg/L is equivalent to ppm (part per million)
milliliter
Liquid Crystal Display

EPA

:

°C

:

°F

:

mg/L

:

mL

:

LCD

:

ABBREVIATIONS

14

• Place the cuvet into the holder and close

the lid.

• Press ZERO and “----” will blink on the display.

• Remove the cuvet from the instrument and

open the cap.

• Add 6 drops of HI 83748B-0 reagent B to the

cuvet.

• After a few seconds the display will show

“-0.0-”. The meter is now zeroed and ready
for measurement

• Reinsert the cuvet into the instrument and

close the lid.

• Replace the cap and shake gently to mix.

x 6

Содержание HI 83748

Страница 1: ...Instruction Manual HI 83748 TARTARIC ACID ISM for wine analysis w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives MAN83748R3 04 06 w w w h a n n a i n s t c o m...

Страница 2: ...N ACCURATE MEASUREMENT 11 MEASUREMENT PROCEDURE 13 BATTERIES REPLACEMENT 16 ACCESSORIES 16 CE DECLARATION OF CONFORMITY 17 WARRANTY 17 HANNA LITERATURE 18 USER NOTES 19 Dear Customer Thank you for cho...

Страница 3: ...s Tissue for wiping cuvets Instruction manual Instrument Quality Certificate Rigid carrying case Note save all packing material until you are sure that the instrument works correctly Any defective ite...

Страница 4: ...n poly phenols and tartaric acid in red and proteins in white wine This explains why as pigment polymerization occurs the holding capacity of tartaric acid diminishes resulting in delayed casse The su...

Страница 5: ...ccurate The figure explains these definitions In a laboratory using a standard solution of 2 0 g L tartaric acid and a representative lot of reagent an operatorobtainedwithasingleinstrumentastandard d...

Страница 6: ...the substance as the other factors are known Photometric chemical analysis is based on the possibility to develop an absorbing compound from a specific chemical reaction between sample and reagents G...

Страница 7: ...zeroing cuvets are optically identical to provide the same measurement conditions Whenever possible use the same cuvet for both It is necessary that the surface of the cuvet is clean and not scratche...

Страница 8: ...READ key to perform measurement 8 Power Socket 12V to 20V DC 2 5 Watt FUNCTIONAL DESCRIPTION FRONT REAR INSTRUMENT DESCRIPTION 13 Turn the instrument on by pressing ON OFF MEASUREMENT PROCEDURE When t...

Страница 9: ...ng a zero measurement The light intensity is automatically re adjusted auto calibration features if necessary 12 Proper use of the dropper a to get good reproducible results tap the dropper on the tab...

Страница 10: ...ries and restart the meter Light low there is not enough light to perform a measurement Please check the preparation of the zero cuvet 11 The instructions listed below should be carefully followed dur...

Отзывы: