ERROR MESSAGES
The meter has lost its configuration. Contact your dealer or the nearest
Hanna Customer Service Center.
“Light high”: there is too much light to perform a measurement.
Please check the preparation of the zero cuvet.
“No Light”: the lamp is not working because of a malfunction. Contact
your dealer or the nearest Hanna Customer Service Center.
a) on zero reading:
a) on zero reading:
a) on zero reading:
a) on zero reading:
a) on zero reading:
A flashing value of the maximum concentration indicates an over
range condition. The concentration of the sample is beyond the
programmed range: dilute the sample and measure again.
The sample absorbs less light than the zero reference. Check the pro-
cedure and make sure you use the same cuvet for reference (zero)
and measurement.
“Inverted”: the sample and the zero cuvet are inverted.
b) on sample reading:
b) on sample reading:
b) on sample reading:
b) on sample reading:
b) on sample reading:
Indicates that the batteries are dead and must be replaced. After this
message appears, the instrument is switched off. Change the batteries
and restart the meter.
“Light low”: there is not enough light to perform a measurement.
Please check the preparation of the zero cuvet.
11
The instructions listed below should be carefully followed during testing to ensure best accuracy.
• For dosing the wine sample, we recommend to use
the supplied Hanna HI 731340 automatic pipette.
For a correct use of the Hanna automatic pipette,
please follow the related Instruction Sheet.
• In order to avoid reagent leaking and to obtain
more accurate measurements, it is recommended
to close the cuvet first with the supplied HDPE
plastic stopper
and then with the black
cap.
• In order to measure exactly 5 mL of reagent with
the 5 mL syringe:
(a) push the plunger completely into the syringe and insert the tip into the reagent bottle.
(b) pull the plunger up until the lower edge of the seal is exactly on the 5 mL mark.
(c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are
hanging on the tip of the syringe, if so eliminate them. Then, keeping the syringe in
vertical position above the cuvet, push the plunger completely down into the syringe. Now
the exact amount of 5 mL has been added to the cuvet.
GENERAL TIPS FOR AN ACCURATE MEASUREMENT
Hanna
automatic pipette
(a)
(b)
(c)