Instruction Manual
HI8314
Portable pH/mV/ºC meter
www.hannainst.com
Hanna Instruments reserves the right to modify the design,
construction or appearance of its products without advance notice.
Dear Customer,
Thank you for choosing a Hanna Instruments product. This
manual will provide you with the necessary information for the
correct operation of the meter. Please read it carefully before
using the meter. If you need additional technical information,
do not hesitate to e-mail us at
.
The instrument is in compliance with CE directives.
Remove the instrument from the packing material and examine
it carefully to make sure that no damage has occurred during
shipment. If there is any damage, please contact your local
Hanna Instruments Office.
The meter is supplied complete with:
•
HI1217D
pH electrode
• calibration and cleaning solution sachets
• 9 V battery, calibration screwdriver and instructions
• quality certificate
Note:
Conserve all packing material until the instrument
has been observed to function correctly. Any defective item
must be returned in its original packing.
PRELIMINARY EXAMINATION
GENERAL DESCRIPTION
HI8314
is an hand-held pH/ mV/ temperature meter designed
to be rugged, easy-to-use, reliable and practical.
It is ideal for education and for measurements on the field.
The pH, mV and temperature ranges are easily selected using
the rubber keyboard on the front panel.
Calibration procedure is very simple with clear help messages
displayed on LCD and adjustments are easily made with two
trimmers on the front panel.
The built-in temperature sensor in the pH electrode allows
automatic temperature compensation of pH readings.
Moreover,
HI8314
features a low battery indicator that alerts
the user when the battery needs to be replaced.
FUNCTIONAL DESCRIPTION
SPECIFICATIONS
1) DIN connector for pH or ORP electrode
2) Liquid Crystal Display (LCD)
3)
pH
key, to display the pH value
4)
mV
key, to display the mV (ORP) readings when using an
ORP electrode or the mV equivalent to pH values when
using a pH electrode
5)
TEMP
key, to display the temperature measurement or
to select temperature unit while temperature is displayed
6)
OFFSET
trimmer
for OFFSET calibration of pH
7)
SLOPE
trimmer for SLOPE calibration of pH
8)
CAL
key, to enter pH calibration
9)
ON/OFF
key, to turn the meter ON and OFF
DISPLAY CODE GUIDE
°C
or
°F
symbol indicates the meter is in temperature mode.
pH
symbol indicates the meter is in pH mode.
mV
symbol indicates the meter is in mV mode.
Range
0.00 to 14.00 pH
±1999 mV
0.0 to 100.0 °C/ 32.0 to 212.0 °F
Resolution
0.01 pH/ 1 mV/ 0.1 °C (0.1 °F)
Accuracy
(@20
°
C/ 68
°
F)
±0.01 pH
±1 mV
±0.4 °C/ ±0.8 °F
(excluding probe error)
pH Calibration
Manual, 2 point, through trimmers
Offset Calibration
±1 pH
Slope Calibration
80 to 110%
Temperature
Compensation
Automatic, 0 to 100 °C (32 to 212 °F)
Electrode (*)
HI1217D (included)
Battery Type
1 x 9V alkaline
Battery Life
450 hours of continuous use
Environment
0 to 50 °C (32 to 122 °F)
RH max 95% non-condensing
Auto-Shut off
after 8’ of non-use
Dimensions
145 x 80 x 36 mm (5.7 x 3.1 x 1.4’’)
Weight
230 g (8.1 oz.)
BATTERY REPLACEMENT
This meter is powered by a 9V alkaline battery.
When ‘’
BATT
’’ tag is displayed blinking a
low battery condition is indicated. When the
low battery indication appears, only a few
hours of working time are left. It is recommended to replace
the battery.
When the battery level is too low to ensure reliable measure-
ments, the meter turns off. It is recommended to replace the
battery immediately. Battery replacement must only take place
in a nonhazardous area using a 9V alkaline battery.
Unscrew the three screws on the rear of the meter, remove the
battery compartment cover and replace the 9V battery with a
new one, while observing the correct polarity.
Make sure the battery contacts are tight and secure before
replacing the cover.