38
• Press
Timer
and the display will show the countdown prior to the measurement or, alternatively, wait
for 2 minutes and 30 seconds and press
Read
. When the timer ends the meter will perform the
reading.
• The instrument displays the results in
mg/L of total chlorine
.
Note
: free and total chlorine have to be measured separately with fresh unreacted samples following
the related procedure if both values are requested.
INTERFERENCES
Interference may be caused by: Bromine, Iodine, Ozone, Oxidized forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake the sample for approximately
2 minutes after adding the powder reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
or acidity greater than 150 mg/L CaCO
3
,
the color of the sample may develop only partially, or may rapidly fade. To resolve this, neutralize the
sample with diluted HCl or NaOH.
Total Chlorine