126
• Add the content of one packet of HI 93751-0
Indicator reagent.
• Replace the cap and invert gently to mix
for 1 minute (about 30 inversions).
• Reinsert the cuvette into the instrument.
• Press
Timer
and the display will show the countdown prior to the measurement or, alternatively, wait
for 5 minutes and press
Read
. When the timer ends the meter will perform the reading.
• The instrument displays concentration in
mg/L of Sulfate (SO
4
2
¯)
.
INTERFERENCES
Interferences may be caused by:
Calcium (as CaCO
3
) above 20000 mg/L
Chloride (as Cl¯) above 40000 mg/L
Magnesium (as MgCO
3
) above 10000 mg/L
Silica (as SiO
2
) above 500 mg/L
Color or suspended matter in large amounts will interfere: suspended matter should be removed by previous
filtration.
Organic matter in large amounts may impede the precipitation of barium sulfate.