When the battery becomes weak, a "

V

" will be displayed to warn the

user that only a few hours of working life are left.
A low battery can result in unreliable measurements. It is recom-
mended to replace it immediately.
Battery replacement must only take place in a non-hazardous area,
and using an alkaline 9V battery.
Slide off the battery compartment cover (see figure below) and replace
the 9V battery with a new one.
Make sure the battery contacts are tight and secure before replacing
the cover.

9 VOLT BATTERY

SLIDE

OFF

HI 933000

BATTERY REPLACEMENT

After every series of measurements, rinse
the probe with tap water.
If a more thorough cleaning is required,
remove the PVC sleeve and clean the
probe with a cloth or a non-abrasive
detergent. When reinserting the sleeve
onto the probe, be sure that the sleeve
is in the right direction with the holes
towards the cable end.
After cleaning the probe, re-calibrate
the instrument.
The probe body is made of PVC. For this
reason it must never come into close
contact with a heat source.
If the probe is exposed to high tempera-
tures (above 50°C/122°F), the rings
might become loose or detached, result-
ing in a serious impairment of the probe.
In such cases, the probe has to be
replaced.

PROBE MAINTENANCE

BATTERY

HI 8033
HI 8633
HI 8733
HI 8734

SLIDE

OFF

Содержание HI 8033

Страница 1: ...Instruction Manual HI 8033 HI 933000 Portable Multi Range Conductivity Meters w w w h a n n a i n s t c o m...

Страница 2: ...al information do not hesitate to e mail us at tech hannainst com These instruments are in compliance with the directives All rights are reserved Reproduction in whole or in part is prohibited without...

Страница 3: ...SPECIFICATIONS OF HI 933000 1 Liquid Crystal Display 2 Rotary switch 3 Knob for manual temperature compensation 4 Calibration knob mS Conductivity meter HI 8033 ins rum ns C k S 0 199 9 S 0 1999 S 0...

Страница 4: ...making sure that the holes on the probe are completely submerged If possible use plastic beakers to minimize any EMC interference H I 7 0 3 0 C CALIBRATION Each meter is supplied complete with a 9V ba...

Страница 5: ...440 86 81300 113800 5096 27 80 6 13370 1467 87 83000 115700 5190 28 82 4 13620 1494 89 84900 117700 5286 29 84 2 13870 1521 90 86300 119700 5383 30 86 14120 1548 92 88200 121800 5479 31 87 8 14370 157...

Страница 6: ...EMENT After every series of measurements rinse the probe with tap water If a more thorough cleaning is required remove the PVC sleeve and clean the probe with a cloth or a non abrasive detergent When...

Страница 7: ...80000 S cm 230 mL bottle HI 7035L 111800 S cm 500 mL bottle HI 7035M 111800 S cm 230 mL bottle HI 7039L 5000 S cm 500 mL bottle HI 7039M 5000 S cm 230 mL bottle HI 7032L 1382 ppm mg L 500 mL bottle H...

Страница 8: ...discharges Avoidtouchingtheseringsatalltimes Duringcalibrationoftheinstruments ESDwriststrapsshouldbeworntoavoidpossibledamagetotheprobe by electrostatic discharge Use plastic beakers to minimize any...

Страница 9: ...2 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 01...

Отзывы: