background image

5

4

1) Liquid Crystal Display
2) ON/OFF key
3) Measurement range selection keys
4) Probe connector

Range

0.0 to 199.9 / 0 to 1999 µS/cm

0.00 to 19.99 / 0.0 to 199.9 mS/cm

Resolution

0.1 / 1  µS/cm ; 0.01 / 0.1 mS/cm

Accuracy

 (@ 20°C / 68°F)

±1% FS (excluding probe error)

Typical EMC Deviation

±2% FS

Calibration

Manual, 1 point through trimmer

Temperature

Automatic, 10 to 40°C (50 to 104°F)

Compensation

with ß = 2%/°C

Probe

 (

included

)

HI 76302W

 with 1 m (3.3') cable

Environment

0 to 50°C (32 to 122°F); RH max 95%

Battery Type

1 x 9 V alkaline

Battery Life

Approx. 100 hours of continuous use

Dimensions

143 x 80 x 38 mm (5.6 x 3.2 x1.5")

Weight

360 g (13 oz.)

FUNCTIONAL DESCRIPTION &

SPECIFICATIONS OF HI 933000

1) Liquid Crystal Display
2) Rotary switch
3) Knob for manual temperature compensation
4) Calibration knob

mS

Conductivity meter

HI 8033

ins rum n s

°C

k%

µ

S

0-199,9

µ

S

0-1999

µ

S

0-19,990

salinity
0-19,990

ppm
CaCO

3

OFF

20

30

10

40

0

50

5

5

0

10

10

15

15

+

-

1

2

3

4

FUNCTIONAL DESCRIPTION &

SPECIFICATIONS OF HI 8033

Range

0.0 to 199.9 ; 0 to 1999 µS/cm

0.00 to 19.99 mS/cm

0 to 19990 ppm

Resolution

0.1 ; 1 µS/cm / 0.01 mS/cm / 10 ppm

Accuracy

 (@20°C/68°F)

±1% FS (excluding probe error)

Typical EMC Deviation

±2 % FS

Calibration

Manual, 1 point

Temperature

Manual, 0 to 50°C (32 to 122°F)

Compensation

with ß = 2%/°C

Probe

 (

included

)

HI 76301W 

with 1 m (3.3') cable

Environment

0 to 50°C (32 to 122°F); RH max 95%

Battery Type

1 x 9 V alkaline

Battery Life

Approx. 100 hours of continuous use

Dimensions

185 x 82 x 47 mm (7.3 x 3.2 x1.9")

Weight

270 g (9.5 oz.)

Содержание HI 8033

Страница 1: ...Instruction Manual HI 8033 HI 933000 Portable Multi Range Conductivity Meters w w w h a n n a i n s t c o m...

Страница 2: ...al information do not hesitate to e mail us at tech hannainst com These instruments are in compliance with the directives All rights are reserved Reproduction in whole or in part is prohibited without...

Страница 3: ...SPECIFICATIONS OF HI 933000 1 Liquid Crystal Display 2 Rotary switch 3 Knob for manual temperature compensation 4 Calibration knob mS Conductivity meter HI 8033 ins rum ns C k S 0 199 9 S 0 1999 S 0...

Страница 4: ...making sure that the holes on the probe are completely submerged If possible use plastic beakers to minimize any EMC interference H I 7 0 3 0 C CALIBRATION Each meter is supplied complete with a 9V ba...

Страница 5: ...440 86 81300 113800 5096 27 80 6 13370 1467 87 83000 115700 5190 28 82 4 13620 1494 89 84900 117700 5286 29 84 2 13870 1521 90 86300 119700 5383 30 86 14120 1548 92 88200 121800 5479 31 87 8 14370 157...

Страница 6: ...EMENT After every series of measurements rinse the probe with tap water If a more thorough cleaning is required remove the PVC sleeve and clean the probe with a cloth or a non abrasive detergent When...

Страница 7: ...80000 S cm 230 mL bottle HI 7035L 111800 S cm 500 mL bottle HI 7035M 111800 S cm 230 mL bottle HI 7039L 5000 S cm 500 mL bottle HI 7039M 5000 S cm 230 mL bottle HI 7032L 1382 ppm mg L 500 mL bottle H...

Страница 8: ...discharges Avoidtouchingtheseringsatalltimes Duringcalibrationoftheinstruments ESDwriststrapsshouldbeworntoavoidpossibledamagetotheprobe by electrostatic discharge Use plastic beakers to minimize any...

Страница 9: ...2 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 01...