2
HI 4004 Calcium Half-cell
HI 4004 Calcium Half-cell
HI 4004 Calcium Half-cell
HI 4004 Calcium Half-cell
HI 4004 Calcium Half-cell
HI 4104 Calcium Combination Electrode
HI 4104 Calcium Combination Electrode
HI 4104 Calcium Combination Electrode
HI 4104 Calcium Combination Electrode
HI 4104 Calcium Combination Electrode
I.
I.
I.
I.
I. Introduction:
Introduction:
Introduction:
Introduction:
Introduction:
The Hanna HI 4004 and HI 4104 are ion selective elec-
trodes designed for the measurement of calcium ions in
aqueous solutions. They utilize a replaceable sensing mod-
ule that contains an organic polymer membrane that is
sensitive to calcium ions. The HI 4004 is a half-cell elec-
trode that requires a separate reference. The HI 4104 is a
combination ion selective electrode.
IIIIIIIIII. Specifications
Specifications
Specifications
Specifications
Specifications
Type:
PVC membrane with
organic ion exchanger
Ion measured:
Calcium (Ca
2+
)
Measurement range:
1.0 M to 3 X 10
-6
M
40080 to 0.12 ppm
Interference:
Organic solvents and cationic detergents must be absent.
Ratio of interfering ion to Ca
2+
must be less than the
ratio indicated below:
15000 for Na
+
sodium
7000
for Mg
3+
magnesium
700
for Ni
2+
nickle
300
for Fe
2+
ferrous
250
for Al
3+
aluminum
200
for NH
4
+
ammonium
35
for Cu
2+
cupric
0.001
for Pb
2+-
lead
Operating Temperature:
0-40
°
C
Operating pH:
4 to 10 pH (see Section
XIII)
Dimensions:
12 mm (OD) X 120 mm
nominal insertion
(0.47
”
X 4.72
”
)
Connection:
BNC