18
L O G G I N G F U N C T I O N
Up to 50 LOG samples can be stored into memory.
LOGGING THE CURRENT DATA
To store the current reading into memory press
LOG
while the instrument is in
measurement mode.
The instrument will display the current date (mm.dd) on the primary LCD, the
record number on the secondary LCD and “LOG” tag will blink for a few seconds
(see example below: record No. 5 dated September 14):
If there are less than 6 memory locations remaining, the record number and
“Lo” message will blink to alert the user.
LOG
LOG
CAL
• The second mV value.
• The Ion calibration slope (mV/decade).
• The instrument ID.
Note:
• If a one-point calibration is performed after a two-point calibration,
the instrument will keep the old slope.
•
If “no bUF” message appears on the LCD, the instrument informs
you that calibration was performed in only one point.
•
Press
GLP
at any moment and the instrument will return to
measurement mode.
•
If calibration has not been performed, the instrument displays
“no CAL” message blinking.
CAL
BUF
2
mV
CAL
mV